To Remove And Replace The Interface And Gpib Board - Agilent Technologies 54830 Series Service Manual

Table of Contents

Advertisement

Chapter 6: Replacing Assemblies

To remove and replace the interface and GPIB board

To remove and replace the interface and GPIB board

Use this procedure to remove and replace the interface and GPIB board, A21. When necessary,
refer to other removal procedures.
Disconnect the power cable and remove the cover.
1
Remove the keyboard cable (W12) and the acquisition board cable (W3) from the
2
interface and GPIB board.
Remove the Torx T10 screw that secures the interface and GPIB board to the rear of
3
the chassis.
Pull the board up to disengage it from the motherboard, then lift up and out of the
4
chassis.
To replace the board, reverse the removal procedure.
5
Figure 6-23
Removing the Interface and GPIB Board
6–24
W12 Cable
W3 Cable
T10 Torx Screw
Interface & GPIB
Board(A21)