Download Print this page

Jadever JIF-2002 Manual page 65

Weighing indicator & controller

Advertisement

※OPEN COLLECTOR TYPE
Maximum Voltage:30V
Maximum Current:24mA
※Please add a pull-up resistance if connected to a TTL LOGIC.
6-5 ANALOG OUTPUT OP-05
◎ Analog Output
F 500 Analog Output Data
F 501 Output Mode
Loss-in-weight
F 502
Absolute Value
Output current
F 503
when display ZERO
Output current at
F 504
Full Capacity
Output Volt when
F 505
display ZERO
Output Volt at Full
F 506
Capacity
OP-5 OUTPUT 4 ~ 20 mA Specifications
Output Level
Resolution
Temperature
Coefficient
Maximum
If you add a 250 Ω resistor , the output will be 1V to 5V (4~20mA)
Output 4~20 mA
Same as display 2Gross data
3NET data
Not read Absolute Value
1 JIF-2002 reads Absolute Value
0.0mA through 9.99mA (Initial 0.40mA)
0.0mA through 9.99mA (Initial 20.0mA)
-2.5V through +59.9V (Initial 00.0 V)
-2.5V through +59.9V (Initial 10.0 V)
4~20 mA effective range. Output range is
approximately 2 to 22 mA
More than 1 / 1000
±(0.015 % / ℃ of rdg + 0.01mA) /℃
500Ω Maximum
JIF-2002 WEIGHING INDICATOR&CONTROLLER
Output 0~+10 V
65

Advertisement

loading