Allen® Advance Table
User Manual
Product No. A-71100-US, A-71100-CA,
A-71101-EU, A-71101-AUS,
A-71101-UK, A-71101-SA,
A-71100-BR, A-71101-PT
A-71101-IND, and A-71101-THA
D-720686-A6

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A-71100-US and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for allen A-71100-US

  • Page 1 Allen® Advance Table User Manual Product No. A-71100-US, A-71100-CA, A-71101-EU, A-71101-AUS, A-71101-UK, A-71101-SA, A-71100-BR, A-71101-PT A-71101-IND, and A-71101-THA D-720686-A6...
  • Page 3 The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of Allen Medical. This manual (D-720686) was originally released and supplied in English. For a list of available translations, contact Allen Medical Technical Support. Sixth Edition, February 2018 First Printing, 2013 Product may be covered by one or more patents.
  • Page 4 Quick View™ is a trademark of Hill-Rom Services, Inc. The information contained in this manual is subject to change without notice. Allen Medical makes no commitment to update or keep current, the information contained in this manual. Allen Medical reserves the right to make changes without notice in design, specifications, and models.
  • Page 5: Table Of Contents

    Air Bladder and Pump ........31 Potential Equalization Conductor ......32 Allen® Advance Table User Manual (D-720686-A6)
  • Page 6 Discard the Unit ..........66 Allen® Advance Table User Manual (D-720686-A6)
  • Page 7 Specifications ........... . 74 Allen® Advance Table User Manual (D-720686-A6)
  • Page 8 NOTES: Allen® Advance Table User Manual (D-720686-A6)
  • Page 9: Intended Use

    Initial version of the Allen® Advance Table. Introduction of Lateral-to-Prone Assembly (see page 13) a. The A2 model of Allen® Advance Table is not available in all countries. Contact your local customer service representative for more information. Allen® Advance Table User Manual (D-720686-A6)
  • Page 10: Symbol Identification

    August 2017 may not be CE marked. Check individual product for details.See page 13. Medical Electrical Equipment Evaluated in accordance with UL60601-1, EN 60601-1, IEC 60601-1, IEC 60601-2-46, and IEC 60601-1-2. Type B applied part. Allen® Advance Table User Manual (D-720686-A6)
  • Page 11 Shows the maximum patient weight for the Support Tops. The Table is not suitable for use in oxygen- enriched environments or where there is a flammable anesthetics mixture with air, oxygen, or nitrous oxide. Battery Charge Level indicator. Allen® Advance Table User Manual (D-720686-A6)
  • Page 12 Shows the deflate control for the air bladder and the connection point for the air pump. Shows that a crushing hazard is present. Make sure that your body is not in an area where a crushing hazard is present. Allen® Advance Table User Manual (D-720686-A6)
  • Page 13 Identifies the controls in the area as column height controls. This is not a control. This symbol is not on all pendants. Up control—raises both columns at the same time. Down control—lowers both columns at the same time. Allen® Advance Table User Manual (D-720686-A6)
  • Page 14 Key control—when pressed along with the unlock control, unlocks the brakes. Lateral surface down—lowers the patient’s legs. This control is not on all pendants. Lateral surface up—raises the patient’s legs. This control is not on all pendants. Allen® Advance Table User Manual (D-720686-A6)
  • Page 15 Manual Flip indicator—is orange when the Table is in Flip mode (the top can be turned manually). Prone Support Set Identifies the chest pad. Identifies the hip pads. Identifies the right thigh pad. Identifies the left thigh pad. Allen® Advance Table User Manual (D-720686-A6)
  • Page 16: Safety Warnings And Cautions

    • No modification of this equipment is allowed. • Patient support tops and accessories from Allen Medical are designed to industry standard dimensions. Before you use products that are not from Allen Medical, you must verify with the product manufacturer that the product can be used safely with the Table.
  • Page 17 • Before you use Allen® Patient Support Tops with a table not manufactured by Allen Medical, you must verify with the product manufacturer that the support tops can be used safely with their product. • There are no known contraindications for use of the Table.
  • Page 18 Do not use tape on the Table. Adhesive residue could affect the functions of the Table or the support tops. However, you may use tape on the lateral top. Use an alcohol wipe to remove adhesive residue. • Keep this manual available for future reference. Allen® Advance Table User Manual (D-720686-A6)
  • Page 19: Quick View™ List Of Features

    Page Item Feature Page Power cord outlet Cervical traction rope insertion On/Off switch Patient support top with pad H-bracket Emergency brake release Cross pin Collapsible cross tube Bowtie Cross pin and H- bracket storage Allen® Advance Table User Manual (D-720686-A6)
  • Page 20: View 2

    Item Feature Page Item Feature Page Power indicator Flip handle 42, 47 Pendant cord holder 60 Pendant holder Pendant connector Pendant Cervical traction Power port for lateral release Display (status Battery charge level indicator) Allen® Advance Table User Manual (D-720686-A6)
  • Page 21: Newer Operating Tables)

    Air bladder (stored in lateral support pad) Rail support top with prone support set Components of the Lateral-to-Prone configuration manufactured prior to August 2017 may not be CE marked. Check individual product for details. Allen® Advance Table User Manual (D-720686-A6)
  • Page 22: Controls And Indicators

    Manual Flip handle—when released, the flip feature can be used. • Cervical Traction release—when pressed, the traction assembly moves out from the column. • Lateral Top Power Point—this supplies power and the control signal to the Lateral Top. Allen® Advance Table User Manual (D-720686-A6)
  • Page 23 • Lateral Top Control (on A2 and newer models)— lowers the leg section of Model A1 Model A2 a lateral top up to a 22° break. • Electronic Brake lock/unlock—locks and unlocks the brakes. Allen® Advance Table User Manual (D-720686-A6)
  • Page 24 You will find the Retract/Extend handle at the center of the cross tube. Lift the Retract/Extend Unlocked: handle to adjust the overall length of the table from 69" to 109" (175.3 cm to 276.9 cm). Lower the handle to lock the cross tube in position. Locked: Allen® Advance Table User Manual (D-720686-A6)
  • Page 25: Setup And Use Instructions

    “beep” from the Table when you try to use the pendant controls, plug the Table into an applicable power outlet to charge the batteries. To keep the batteries charged, plug the Table into an outlet nightly and when you store the Table. Allen® Advance Table User Manual (D-720686-A6)
  • Page 26: Potential Equalization Conductor

    Unlocked indicator is orange. During a loss of power, if the brakes are locked and it is necessary to move the Table, use the Emergency Brake Release on both columns to manually unlock the brakes. Allen® Advance Table User Manual (D-720686-A6)
  • Page 27: Manually Release The Brakes

    Lift the cross tube handle to unlock the cross tube. With a person at each end of the Table, adjust the length as necessary: • Retract—push the columns toward each other. • Extend—pull the columns away from each other. Allen® Advance Table User Manual (D-720686-A6)
  • Page 28: Install The H-Brackets And A Patient Support Top

    Table, do Step a through Step i. Otherwise, go to Step 5 on page 22. Hold the H-bracket so that the open end is down, the J ends are toward you, and the bracket is angled toward you. Allen® Advance Table User Manual (D-720686-A6)
  • Page 29 Make sure the drop lock on the end of the cross pin moves to the side to hold the H-bracket in position. Repeat these steps to install an H- bracket on the other column. Allen® Advance Table User Manual (D-720686-A6)
  • Page 30 Locked position. You do not need to press the red Safety lever. Determine which hole location on the H-bracket to install the support top. NOTE: The hole locations are numbered on the side of the H-brackets. Allen® Advance Table User Manual (D-720686-A6)
  • Page 31: Remove A Support Top And/Or Bracket

    Never remove a cross pin from a support top while a patient is on the support top. Patient fall or injury could occur. Make sure the patient has been moved from the support top. Do the applicable steps of the install procedure in reverse order. Allen® Advance Table User Manual (D-720686-A6)
  • Page 32: Install The Advance Pinless H-Bracket

    Locked indicator on the Table display is green). Do the steps that follow to open the mounting latch on both sides of the Allen® Advance Pinless H-Bracket: Press the release button on the mounting latch. Move the release button away from the center of the Advance Pinless H-Bracket to put the mounting latch in the open position.
  • Page 33 Make sure that the red safety indicator is not visible d. If necessary, repeat Step a through Step c to install an Advance Pinless H-Bracket on the other column. Allen® Advance Table User Manual (D-720686-A6)
  • Page 34 Push the Flip handle up to the Locked position. You do not need to press the red Safety lever. If necessary, repeat Step a through Step d to install a second pair of Advance Pinless H-Brackets on the bowties. Allen® Advance Table User Manual (D-720686-A6)
  • Page 35: Install A Patient Support On The Advance Pinless H-Bracket

    Install a Patient Support on the Advance Pinless H-Bracket WARNING: Do not use the Allen® Lateral Top with the Advance Pinless H-Bracket. Injury or equipment damage can occur. Put the tilt control release in the unlocked position. Each tilt control release will operate the other tilt control release.
  • Page 36: Install The Lateral Bracket

    Make sure the drop lock on the end of the cross pin moves to the side to hold the H-bracket in position. d. Repeat Step a through Step c at the other end of the table. Allen® Advance Table User Manual (D-720686-A6)
  • Page 37 Push the Flip handle up to the Locked position. You do not need to press the red Safety lever. Allen® Advance Table User Manual (D-720686-A6)
  • Page 38: Install The Lateral Top

    The foot end of the lateral top is not designed to support the weight of the patient’s torso. Personal injury, patient fall, and/or equipment damage can occur. Position the head end of the lateral top at the head end of the Allen® Advance Table. •...
  • Page 39: Air Bladder And Pump

    To deflate the air bladder, activate the deflate control on the air manifold at the head end of the lateral top. Allen® Advance Table User Manual (D-720686-A6)
  • Page 40: Potential Equalization Conductor

    Lay the closed part of the sling over the support top rails. Connect the buckles. Adjust the straps as necessary for the amount of slack you want in the sling. Allen® Advance Table User Manual (D-720686-A6)
  • Page 41: Hip And Thigh Supports

    Slide the blue clamp to the locked position. When locked, only the green area will show on the pin. NOTE: Look under the rail to make sure the clamp is correctly engaged on the locking posts. Allen® Advance Table User Manual (D-720686-A6)
  • Page 42: Chest Support

    The black section should not be covered. NOTE: Allen® Support Pads are designed to work with Allen® Ultra Comfort ™ Covers only. To remove a support from the Table, slide the blue clamp to the unlocked position.
  • Page 43: Head Plate

    Install an Allen® Ultra Comfort™ Cover on to the support. Make sure that— • The foam side of the cover is on the outside. • The cover completely covers the blue material of the support. NOTE: Allen® Support Pads are designed to work with Allen® Ultra Comfort™...
  • Page 44: Pre-Use Check

    Before each use of the Table, make sure all necessary accessories are attached and the Table is ready for use. If any of the controls or indicators do not operate correctly, do not use the Table. Contact Allen Medical or your local distributor.
  • Page 45: Adjust The Height Of The Table

    Press and hold the applicable Raise or Lower control until the Table is at the correct height. NOTE: Near the bottom, on the outside of the columns, there are numbers to help identify the height level. Allen® Advance Table User Manual (D-720686-A6)
  • Page 46: Change The Height Of The Advance Pinless H-Bracket

    Failure to do so could cause patient fall, injury, or equipment damage. Press and hold the lift trigger to release the mechanism. Each lift trigger will operate the other lift trigger. Allen® Advance Table User Manual (D-720686-A6)
  • Page 47: Put The Table In A Trendelenburg Position

    NOTE: We recommend that you use the two large security straps to secure the patient to the support top. • The Table is free of objects that could interfere with the adjustment. Allen® Advance Table User Manual (D-720686-A6)
  • Page 48: Tilt The Support Top

    Make sure that— • The brakes are locked (the Locked indicator on the Table display is green). • The patient is secured to the support top and is fully prepared for the adjustment. Allen® Advance Table User Manual (D-720686-A6)
  • Page 49: Level The Table And The Support Top

    Before you adjust the Table, make sure that— • The brakes are locked (the Locked indicator on the Table display is green). • The patient is secured to the support top and is fully prepared for the adjustment. Allen® Advance Table User Manual (D-720686-A6)
  • Page 50: Rotate The Support Top 90

    Table is free of objects that could interfere with the adjustment, and all loose or unsupported objects have been secured. Otherwise, patient fall, injury, or equipment damage could occur. Make sure that— Allen® Advance Table User Manual (D-720686-A6)
  • Page 51 H-bracket in position. d. Repeat Step a through Step c at the other end of the table. Install the rail top with the prone support set as follows: Allen® Advance Table User Manual (D-720686-A6)
  • Page 52 35 for placement procedures. Use the Drawsheet to pull the patient towards the rail top with the prone support set. Secure the Drawsheet to the rails. Refer to the instructions included with the Drawsheet. Allen® Advance Table User Manual (D-720686-A6)
  • Page 53 Have a caregiver gently lift up on the rail top with prone support set to relieve the weight from the flip handle mechanism. Relieving the weight will reduce the risk that the mechanism will bind and suddenly release. Allen® Advance Table User Manual (D-720686-A6)
  • Page 54 The Pendant indicator on the display turns green to show that you can use the pendant controls. 11. Make sure the support tops are locked in position. If the tops did not lock, return to Step 8. Allen® Advance Table User Manual (D-720686-A6)
  • Page 55: Flip The Support Top 180

    Table is free of objects that could interfere with the adjustment, and all loose or unsupported objects have been secured. Otherwise, patient fall, injury, or equipment damage could occur. Make sure that— Allen® Advance Table User Manual (D-720686-A6)
  • Page 56 To do so could cause patient fall, injury, or equipment damage. d. Make sure the drop lock on the end of the cross pin moves to the side to hold the H-bracket in position. Allen® Advance Table User Manual (D-720686-A6)
  • Page 57 Patient fall or injury could occur. h. At each end of the support top, insert a cross pin into the H-bracket, through the top connector assembly, and through the other end of the H- bracket. Allen® Advance Table User Manual (D-720686-A6)
  • Page 58 Loosely install a minimum of four safety straps evenly spaced and around both support tops. NOTE: We recommend that you use the four Allen® Safety Straps that were included with the support top. Make sure the support tops are in the correct position, and then tighten the safety straps.
  • Page 59 Persons and equipment will not be in the way of the support tops when they get flipped. • To watch the lines so that they don’t get pinched or caught during the flip. • The patient’s head stays in the correct position during the flip. Allen® Advance Table User Manual (D-720686-A6)
  • Page 60 Table is in Flip mode. After the patient has been flipped, and the tops are level (the Tilt Level indicator on the display is green), push the Flip handle up to the Allen® Advance Table User Manual (D-720686-A6)
  • Page 61: Transfer A Patient

    (The Pendant Control indicator on the Table display is green.) • The patient and staff are fully prepared for the transfer. Make sure the Table support top and the surface of the transfer device are at the same height. Allen® Advance Table User Manual (D-720686-A6)
  • Page 62: Access The Cervical Traction Assembly

    The rope will go through the traction assembly and out the left side of the column, near the release. Put the traction rope over and around the pulley. Allen® Advance Table User Manual (D-720686-A6)
  • Page 63 Pull the traction rope from the bowtie down and underneath the bar. NOTE: The bar has an indentation to help keep the rope in position during cervical traction. Allen® Advance Table User Manual (D-720686-A6)
  • Page 64 After use of the traction weight, remove the weight and the rope from the assembly. Press the Cervical Traction Release button, and then push the assembly into the column so that it is in the locked-in position. Allen® Advance Table User Manual (D-720686-A6)
  • Page 65: Setup And Use The Advance Cervical Device

    ACD is secure. Make sure that the ACD is secure on the top side of the bowtie. 4. Push the Cervical Traction Release button. The assembly moves out from the column. Allen® Advance Table User Manual (D-720686-A6)
  • Page 66 See “Change the Cervical Traction Angle” on page 59. WARNING: The maximum weight load for the traction assembly is 50 lb (22.6 kg). To exceed this weight could cause Allen® Advance Table User Manual (D-720686-A6)
  • Page 67: Change The Cervical Traction Angle

    Patient injury can occur. 1. If necessary, fully support and stabilize the patient so that the patient is not in cervical traction. Allen® Advance Table User Manual (D-720686-A6)
  • Page 68: Shut Down The Table

    Table. NOTE: We recommend that you keep the Table plugged in when it is not in use so that the batteries will stay charged. Know and follow your facility’s protocol. Allen® Advance Table User Manual (D-720686-A6)
  • Page 69: Clean And Disinfect

    Do not use iodophors; they will stain the product. CAUTION: If you use a quaternary-type disinfectant, make sure to rinse it off the product immediately. Otherwise, the disinfectant could cause corrosion on painted and plated components. Allen® Advance Table User Manual (D-720686-A6)
  • Page 70: Clean

    When there is visible soil and also between patient use, we recommend that you disinfect the unit with an EPA registered (US only), tuberculocidal, disinfectant such as Morning Mist Neutral Disinfectant. Dilute the disinfectant as shown on the manufacturer's label. Allen® Advance Table User Manual (D-720686-A6)
  • Page 71: Troubleshooting

    Only Allen Medical-authorized persons should service the Table. Service by unauthorized persons could cause patient fall, injury, or equipment damage. If the troubleshooting information below does not fix the problem, contact Allen Medical Technical Support “Service Calls” on page 64. ABLE O INTO ENDANT...
  • Page 72: Service Calls

    SERVICE CALLS When you call Allen Medical about your unit, be prepared to give the serial number from the product identification label. You will find the label on the inside of the head-end base near the on/off switch. PREVENTIVE MAINTENANCE WARNING: Only Allen Medical-authorized persons should service the Table.
  • Page 73: Storage And Handling

    75. If you have the optional accessory cart, store the support tops and additional accessories on the cart. Use the optional accessories cart to store the Allen Advanced Lateral top and brackets. Allen® Advance Table User Manual (D-720686-A6)
  • Page 74: Expected Life

    When correctly maintained, the battery has an expected life of 3 years. DISCARD THE UNIT When the Table has reached the end of its useful life, discard it in accordance with local and federal standards. Recycle as much as possible. Allen® Advance Table User Manual (D-720686-A6)
  • Page 75: Parts Replacement

    Pendant A-71403 Pendant nut C-007239 H-bracket A-71402 Prone support bracket A-71702-P Lateral support bracket A-71702-L Cross pin A-71401 Power cord, EU A-71407 Power cord, UK A-71406 Power cord, US A-71405 Power cord, IND/SA A-71408 Allen® Advance Table User Manual (D-720686-A6)
  • Page 76: Pendant

    Unscrew the nut from the Table connector. Remove the nut from pendant cord, and install the nut on the new pendant cable. Make sure the ribbed end of the nut is toward the connector end of the pendant cable. Allen® Advance Table User Manual (D-720686-A6)
  • Page 77: Replacement

    Table. Pull the fuse holder from the plug housing. NOTE: You may need to use a small screwdriver or a pen to pry the holder from the housing. Remove the fuses from the holder. Allen® Advance Table User Manual (D-720686-A6)
  • Page 78: Replacement

    Install the new fuses in the holder. Fully insert the holder into the plug housing. Plug the power cord into the Table and into an applicable power outlet, and turn the Table on. Make sure all Table functions operate correctly. Allen® Advance Table User Manual (D-720686-A6)
  • Page 79: Accessories

    Arm support, (US) A-70700 C-Flex® head A-70701 C-Flex® base unit positioner system A-70702 C-Flex® clean cape A-70710 Starburst adapter for disposables standard Mayfield or DORO A-70715 Prone mask head A-70720 Flat plate head module module Allen® Advance Table User Manual (D-720686-A6)
  • Page 80 A-71311 Thigh, support with large pad, right A-71312 Sling support A-71316 Head plate A-71400 Rail top A-71404 Cervical traction bar A-71410 Flat top A-71411 Safety straps A-71416 Advance Pinless A-71417 Advance Cervical H-Bracket Device Allen® Advance Table User Manual (D-720686-A6)
  • Page 81 Accessory cart (US) accessory rail clamp (UK) A-71600-EU Accessory cart (EU) A-71600-UK Accessory cart (UK) a. A-70700 includes one each: A-70701, A-70702, A-70310, A-70710, A-70715, and A-70720. b. A-70800 includes one each: A-70801, A-70821, and A-70815. Allen® Advance Table User Manual (D-720686-A6)
  • Page 82: Specifications

    Product Number Description A-71100-XX* Allen® Advance Table, 120 V A-71101-XX* Allen® Advance Table, 230 V A-71705-XX* Allen® Air Pump, 120 V or 230 V * XX is a placeholder for the country code. Dimensions Feature Dimension Width 32.5" (82.6 cm) Column height (measured from the 26.5"...
  • Page 83 120 V - 60 Hz A-71101-XX 230 V - 50 Hz A-71705-US 120 V - 60 Hz A-71705-EU/UK 230 V - 50 Hz Maximum Current Draw Model Range A-71100-XX A-71101-XX A-71705-US 0.24 A A-71705-EU/UK 0.13 A Allen® Advance Table User Manual (D-720686-A6)
  • Page 84 Type B applied part According to Degree Protection against Electrical Shock IEC 60601-2-46 Classification IPX 4 According to Degree of Protection against Dust and Fluid Intrusion Mode of Operation 2 minutes ON/18 minutes OFF Allen® Advance Table User Manual (D-720686-A6)
  • Page 85 Flat Top Positioning supports (such as Flat Top pad pads, C-Flex® Head Positioning System, and disposables) Lateral top Lateral top pad Lateral Drawsheet Lateral-to-prone bracket set Air pump Air pump tube set Air bladder Allen® Advance Table User Manual (D-720686-A6)
  • Page 86 (RF) communications equipment can affect electrical equipment. Medical equipment needs special precautions in regard to electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information supplied in the tables that follow. Allen® Advance Table User Manual (D-720686-A6)
  • Page 87 Harmonic Class A Emissions IEC 61000-3-2 Voltage Complies Fluctuations/ Flicker Emissions IEC 61000-3-3 Allen® Advance Table User Manual (D-720686-A6)
  • Page 88 Magnetic Field be at levels characteristic of a typical IEC 61000-4-8 location in a typical commercial or hospital environment. Note: U is the AC mains voltage prior to application of the test level. Allen® Advance Table User Manual (D-720686-A6)
  • Page 89 If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating or A-71705-XX the A-71100-XX/A-71101-XX Over the frequency range of 150 kHz to 80 MHz, field strength should be less than 3 Allen® Advance Table User Manual (D-720686-A6)
  • Page 90 NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Allen® Advance Table User Manual (D-720686-A6)