Receiving

Moving the Cabinet, Testhead and Accessories to Your Clean Room

Moving the Cabinet, Testhead and Accessories to Your Clean Room
This section covers how to move and unwrap the cabinet, testhead, and accessories in your clean room.
All items are wrapped in double protective vinyl covers.

For the Agilent 4073A Testhead

1. Move the 4073A to the outside entrance of the clean room.
2. Remove the outer vinyl covers, but do not cut the deaeration packaging's vinyl cover.
3. Move the 4073A into the buffer room, then remove the vinyl covers using the following procedure.
CAUTION
You must ensure that no condensation occurs in the testhead. Before removing the deaeration
packaging's vinyl cover, the testhead must adapt to the ambient temperature and humidity. If
condensation does occur in the testhead, the tester may not meet specifications.
• If the testhead has sat at the entrance of the clean room for more than 8 hours, measure the
atmospheric temperature at the entrance, then wait for the corresponding time shown in Table
4-1.
• If the testhead has sat at the entrance of the clean room for less than 8 hours, measure the
atmospheric temperature of the location the testhead sat at prior to being moved to the
entrance, then wait for the corresponding time shown in Table 4-1.
Example:
If the temperature of the air outside is colder than the entrance, and you move the testhead
to the entrance after putting on the cold place for log time, you need to check the
atmospheric temperature at the place, then wait fot the corresponding time shown in Table
4-1.
The corresponding with the waiting time for no-condensation is the following Table 4-1.