ELECTRONIC DIGITAL THREE-POINT
INTERNAL MICROMETERS
Thank you for purchasing the Digital 3-Point Micrometer.
Please read this manual thoroughly before use for proper operation.
FEATURES
Easy to read digital display.
High accuracy 3-point ID measurements.
Includes extension rod (Deep Hole Extension Rod) for measurements in deep measurements.
With constant pressure ratchet mechanism.
SPECIFICATIONS
Resolution : 0.001mm
Probe Material : Carbide tipped
Power : SR44 ( Silver Oxide Battery)
※1x included for testing
Auto-display-off function (approx. 5 minutes.)
Measuring
Maximum
Weight
Model No.
Range (mm)
Error (µm)
(g)
6 −
MCD3385-0608HTN
8
180
8 − 10
MCD3385-0810HTN
180
10 − 12
MCD3385-1012HTN
180
MCD3385-1216HTN
12− 16
230
4
MCD3385-1620HTN
16− 20
230
MCD3385-2025HTN
20− 25
300
MCD3385-2530HTN
25− 30
320
MCD3385-3040HTN
30− 40
420
Dedicated Spanner (2x)
MCD3385-4050HTN
40− 50
510
MCD3385-5063HTN
50− 63
710
MCD3385-6275HTN
62− 75
5
720
MCD3385-7588HTN
75− 88
990
MCD3385-87100HTN
87−100
1010
Dimensions (mm)
Model No.
A
B
C
D
MCD3385-0608HTN
MCD3385-0810HTN
Less than 2
85
100
MCD3385-1012HTN
MCD3385-1216HTN
100
MCD3385-1620HTN
MCD3385-2025HTN
105
MCD3385-2530HTN
43
MCD3385-3040HTN
Less than 1
110
150
MCD3385-4050HTN
MCD3385-5063HTN
MCD3385-6275HTN
127
MCD3385-7588HTN
MCD3385-87100HTN
INSERTING THE BATTERY
1.
Using the battery cover tool, open the battery cover by rotating in the direction
Open
of the arrow.
2.
Insert battery with "+" side outward.
Replace battery cover.
3.
Battery Cover
Use only SR44 type battery.
NOTICE
Micrometer is a precision instrument. For specified performance, use in environment around 20° C and allow sufficient time to acclimatize.
Temperature extremes may affect instrument accuracy and operation.
INSTRUCTION MANUAL
Part Number : MCD3385- ** HTN
PART IDENTIFICATION
Ratchet Thimble
(constant pressure mechanism)
Gauge Tips
LCD Display
... SET
ABS/INC...UNIT
Gauge Head
Rod
... SET
[
] Button
[ ABS/INC ... UNIT ] Button
Battery Cover
Deep Hole Extension Rod
SR44
SR44 (Silver Oxide Battery)
※Battery provided for testing.
Battery Cover Tool
 Please replace with fresh
[ Body : MCD3385-0608HTN〜MCD3385-4050HTN ]
A
B
[ Body : MCD3385-5063HTN〜MCD3385-87100HTN ]
A
B
[ Deep Hole Extension Rod ]
D
PLEASE NOTE
SETTING THE ORIGIN
Set the origin to be used for measurement.
The origin is stored even when power is off, so it does not have to be reset every time
<combined>
power is turned on. When the battery is replaced, the origin will have to be reset.
1.
Prepare instrument and ring gauge by cleaning measurement head and ring gauge
inner diameter with appropriate cleaner.
... SET
2.
Press the [
Press and hold the [ ... SET
3.
on LCD and will start blinking when you release the button.
Confirm that the displayed value matches the dimensions of the ring gauge standard.
If they do not match, use the following procedure to modify.
With "Set" flashing on LCD, a long press on the [
blinking to the readout digits in sequence. A short press on the [
change the value at the cursor. Repeat until correct value has been entered and "Set"
blinks.
4.
Insert the measurement head into the ring gauge and turn the thimble until the
measuring surfaces make contact with the inside of the ring gauge. Continue turning
until ratchet stop "clicks" 1~3 times.
Press the [ ... SET
5.
should match the dimension of the ring gauge standard, and origin setting is
complete.
DEEP HOLE EXTENSION ROD
Use the following procedure and the special spanners provided to attach the deep hole
extension rod.
1.
Place the accessory spanners into the notches above and below the split in the rod
and turn to separate the rod into two parts, the "gauge head" and "display unit."
See Figure 2
【BACK】
2.
Attach the extension rod to the measurement head and then attach to the display
unit. Tighten with accessory spanners.
3.
Re-set the origin using the procedure above -- "Setting the origin"
Attach
Gauge Head
Deep Hole Extension Rod
OPERATION
1.
Remove any grease and dust from the measuring surfaces with a clean cloth, such as a lens cleaning cloth. Make sure no foreign objects
are adhered to the surfaces.
... SET
Press the [
2.
Make sure "ABS" is displayed on the LCD; press the [ ABS/INC ... UNIT ] if required.
3.
4.
Insert the measurement head into object to be measured and turn ratchet thimble until the measuring surfaces make contact with the inside
diameter. Continue turning until ratchet "clicks" 1~3 times and then read the measurement off the LCD display.
INC Mode
Press the [ ABS/INC ... UNIT ] button to reset display to zero while retaining the origin setting in memory.
To return to normal (absolute) measurement mode, press the [ ABS/INC ... UNIT ] button again.
TROUBLESHOOTING
ERROR CONDITION
Measured value is wrong
Gauge head does not retract
Nothing is displayed on LCD
Display is not stable
Display is not clear
※If error continues, or if you have any questions, please contact distributor or place of purchase.
※Please note, manufacturer is unable to respond to inquires or provide service directly. Please contact distributor or place of purchase.
CAUTION
Always remove dirt from measuring surfaces with a lens cloth before use.
Do not turn thimble more than 3mm above upper measurement range.
This is a precision instrument, handle with care. Do not drop or subject to excessive forces.
Keep instrument away from direct sunlight or high temperatures such as in a car, or near a stove or heat source.
Keep instrument away from water and oil.
Please do not disassemble or modify.
Remove dust and cutting chips after use and apply rust preventative oil to metallic surfaces. Please keep out of
reach of children.
When instrument will be idle for one month or more, please remove the battery to prevent damage from battery
fluid leakage.
Please do not use organic solvents to clean the instrument body.
Only use as directed. Improper use may cause accident or injury.
※Prepare a Ring Gauge to use as a reference standard
] button to turn on power.
] button for more than thee seconds, "Set" will display
... SET
] button will move the
... SET
] button will
See Figure 1
] button. The flashing "Set" will disappear. The displayed value
See Figure 3
Figure 3
Attach
...SET
ABS/INC...UNIT
Display Unit
] button to turn on power.
CORRECTIVE ACTION
Check if measuring surfaces are dirty. Clean with lens cloth.
Check to insure origin-point is correct. → refer to section "Setting the origin"
When stored at low temperature, or not used for a while, the internal grease may become stiff.
Allow sufficient time for grease to warm up.
Wearing dust-free gloves, use hand to push in.
Remove and reinsert battery to cycle power.
Replace the battery.
Figure 1
"Click, Click"
Ratchet Thimble
2
Turn
... SET
ABS/INC...UNIT
3
Press
1
Confirm displayed
value matches
ring gauge dimensions.
Figure 2
Rod
Notch
Loosen
Spanner
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp
URL
http://www.niigataseiki.co.jp

Advertisement

loading

Summary of Contents for Niigata seiki MCD3385-0608HTN

  • Page 1 … SET Press the [ ] button to turn on power. [ Body : MCD3385-0608HTN〜MCD3385-4050HTN ] Make sure “ABS” is displayed on the LCD; press the [ ABS/INC … UNIT ] if required. Model No. Insert the measurement head into object to be measured and turn ratchet thimble until the measuring surfaces make contact with the inside MCD3385-0608HTN diameter.
  • Page 2 ※付属の電池はテスト用です。 ご使用 電池蓋開閉スパナ MCD3385-87100HTN 87 - 100 1010  の際は新たにお買い求めください。 ABS/INC…UNIT … 測定部側 深穴測定用延長ロッ ド LCD部側 図3 図2 寸法図 [ 本体: MCD3385-0608HTN 〜 MCD3385-4050HTN ] (単位 : mm) 使 ①測定子を含めた測定部を脱脂し、グラスペーパなどで拭き取って異物が付着していない状態にしてください。 A B C D 品 番 用 方 ②“ …SET ”ボタンを押して、電源を入れてください。 MCD3385-0608HTN 法...

This manual is also suitable for:

Mcd3385-0810htnMcd3385-1012htnMcd3385-1216htnMcd3385-1620htnMcd3385-2025htnMcd3385-2530htn ... Show all