Battery Replacement; Immersion Depth - ThermoProbe TL1 Series Instructions Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

BATTERY REPLACEMENT

WARNING:
Batteries must be changed in Non-hazardous area.
Batteries must be of correct approved type.
Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+) symbol embossed
in the battery case.
New batteries must not be mixed with old batteries. Batteries must not be mixed with batteries of other manufacturers.
Batteries must not be installed with polarity reversed where one cell could charge another cell.
a) Ensure the instrument is in a non-hazardous area.
b) Remove the 2 screws and back cover on the TL1 to access the batteries.
c) Install each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the battery case.
d) For models with AAA batteries, the user must make certain that the rubber spacer remains in place after servicing the
batteries. This will ensure the batteries cannot shift during use.
e) Replace the back cover and 2 screws.
CERTIFIED Batteries are as follows:
Model
TL1-W, TL1-R
TL1-W, TL1-R
TL1-W, TL1-R
TL1-A

IMMERSION DEPTH

For best measurement results, a minimum immersion depth of 4 inches into the temperature source is recommended.
All TL1 models shall maintain a minimum separation between the sensor mount and temperature source to ensure that the
thermometer enclosure does not exceed 40°C.
Figure 1: TL1 immersion depth
09/2012, JK
Manufacturer
Duracell
Panasonic
Gold Peak
Duracell
Type
AAA (LR03) Alkaline
AAA (LR03) Alkaline
AAA (LR03) Alkaline
CR2032 LiMnO
2
Part Number
MN2400
LR03XWA
GP24A
DL2032

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tl1-aTl1-wTl1-r

Table of Contents