Download Print this page

Mesalabs IBP HDM97Pocket Series Safety Instructions And Quick Start Manual

Reference meter

Advertisement

Quick Links

HDM97Pocket Reference Meter Series
Safety Instructions and Quick Start Guide
HDM97
HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ
Safety Instructions and Quick Start Guide
This document describes the major functions of the HDM97Pocket instrument series. The complete manual can
HDM97
HDM97
BL
be found at http://www.ibpmt.com/medical_e/meter_documents.shtm and on the installation CD. For reading
the document Adobe
This document describes the major functions of the HDM97Pocket instrument series. The complete manual
can be found at http://www.ibpmt.com/medical_e/meter_documents.shtm and on the installation CD. For
®
Windows
is included in several language versions on the CD.
reading the document Adobe
®
Microsoft
Windows
Reference Meter Series
Pocket
HDM97
BM&BH
®
PDF it's necessary to have Adobe
®
PDF it's necessary to have Adobe
®
is included in several language versions on the CD.
HDM97
HDM97
BN
BO
®
PDF Reader installed. Adobe
®
PDF Reader installed. Adobe
Display
Display
Reset / tare pressure / adjust (pH)
Reset / Tare pressure / Adjust (pH)
and navigation key "OK"
and navigation key
Special functions and navigation key
Special functions
"next"
and navigation key
Select functions and navigation key
Select functions
"previous"
and navigation key
ON / OFF
ON / OFF
For OFF press 3 sec.
For OFF press 3 sec.
Channel select
Channel select
and navigation key "ESCAPE"
and navigation key
1
HDM97
BP
BQ
®
®
Reader for Microsoft
®
Reader for
"OK"
"next"
"previous"
"ESCAPE"
www.ibpmedical.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the IBP HDM97Pocket Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Mesalabs IBP HDM97Pocket Series

  • Page 1 HDM97Pocket Reference Meter Series Safety Instructions and Quick Start Guide HDM97 Reference Meter Series Pocket HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ Safety Instructions and Quick Start Guide This document describes the major functions of the HDM97Pocket instrument series. The complete manual can HDM97 HDM97 HDM97...
  • Page 2: Safety Instructions

    Safety Instructions For Your and Your Patients Safety Read and Follow Carefully the Safety Instructions Below • Read the entire manual (not only this guide) carefully before using the HDM97Pocket. • Keep the instrument away from unauthorized persons. • Only use the HDM97Pocket as reference instrument to measure Conductivity, Temperature, Pressure, Flow and pH depending to it’s equipped features.
  • Page 3 Access Connectors HDM97BH HDM97BL Flow Pressure HDM97BN HDM97BO Pressure Conductivity Conductivity Temperature Temperature HDM97BP HDM97BQ Pressure Conductivity Conductivity Flow Pressure Temperature Temperature HDM97BQ with Option C (conductivity probe connector) Flow Conductivity Pressure Temperature Depending on the configuration the appliance part consists of a BNC inlet for pH electrode connection, a male connector for pressure transducer inlet, a connector for the flow probe and a directly attached cable which leads to the conductivity / temperature probe or connector for the probe.
  • Page 4: Basic Settings

    Basic Settings BASIC SETTINGS After switching on the instrument immediately press “F” key to invoke the basic settings menu. The parameter to change can be selected using the navigation keys down arrow and up arrow. By pressing the navigation key (the “ok” button) you can adjust the value of each parameter.
  • Page 5 TIMER stopwatch function enabled stopwatch function disabled Then to apply and save setting, to discard this setting. Pressure Hold Test increases the value lowers the value Then to apply and save setting, to discard this setting. Only available with version BP pH enabled pH enabled Then...
  • Page 6 Versions The HDM97Pocket meters are equipped with up to 5 channels, depending on the version. Channel HDM97BL HDM97BH HDM97BN HDM97BO HDM97BP HDM97BQ Conductivity Temperature Timer Pressure Flow Operating The functions of the keys are structured hierarchically from left to right. Channel Select CD/TEMP (Available in versions BN, BO, BP and BQ)
  • Page 7: Timer Function

    (Available only in versino BP, if enabled in the Basic Settings) Selection between compensation 20°C, 25°C, 37°C and measured temperature Reading max. value Reading min. value Returns automatically back to measuring mode after 5 sec. Reset of both the min. And max. Value Return to measuring display No function Flow...
  • Page 8 Flow-through Adapter for Conductivity and Temperature Measurement The adapter is suitable for measuring pressure, temperature and conductivity in flow- though mode. With immersion usage of the probe it is suitable for measuring temperature and conductivity concurrently. The preferred orientation of the adapter is vertical. The axial connector must point down (see picture).

This manual is also suitable for:

Ibp hdm97blIbp hdm97bhIbp hdm97bnIbp hdm97boIbp hdm97bpIbp hdm97bq