1
General Information
This document contains necessary information for the proper installation and use of this
device. In addition to this instruction, be sure to observe all statutory requirements, appli-
cable standards, the additional technical specifications on the accompanying data sheet
(see www.labom.com) as well as the specifications indicated on the type plate.
1.1
General Safety Notes
The installation, set up, service or disassembly of this device must only be done by
trained, qualified personnel using suitable equipment and authorized to do so.
Warning
Media can escape if unsuitable devices are used or if the installation is not
correct.
Danger of severe injury or damage
 Ensure that the device is suitable for the process and undamaged.
1.2
Intended Use
The device is suitable for measuring the filling level as specified in the data sheet.
1.3
Conformity with EU Regulations
The CE-marking on the device certifies its compliance with the applicable EU Directives for
placing products on the market within the European Union.
2
Transportation and Storage
Store and transport the device only under clean and dry conditions preferably in the origi-
nal packaging. Avoid exposure to shocks and excessive vibrations.
Permissible storage temperature:
3
Installation and Commissioning
Ensure that the device is suitable for the intended application with respect to pressure
range, overpressure limit, media compatibility, temperature range and process connection.
3.1
Mechanical Installation
For installation use only approved sleeve clamps and adapters. Installation in non-Labom
systems can result in incorrect function and leakage.
The thread of the devices must be electrically connected with the threaded sleeve and the
metallic container.
Insulating seals of Teflon or paper are not suitable for correct, error-free operation.
The tightening torque must be between 5 and 10 Nm when screwing the device into the
sleeve.
With non-metallic containers, a threaded sleeve with a diameter of at least 28 mm must be
used.
If the device is opened there is a risk of influencing the signals by touching the electronic
connections. This can be avoided by switching off the supply voltage or disconnecting the
signal circuit.
BA_079_2019-05_10.00 Limit switch LIMES
-20...70 °C
Page 2/5