Problem
The sound sensor
lights orange and "
symbol appears.
WHEEZE
NO WHEEZE
The sound sensor
does NOT flash or
light orange but "
"
symbol appears.
flashes
lights
No power.
Nothing appears.
Results change when
taking measurements
repeatedly.
If you are not able to solve the problem, contact your OMRON retail outlet or distributor.
Possible cause
There is an external noise
"
around the device.
An internal error
occurred.
Batteries are low.
Batteries are depleted or
battery polarities are not
properly aligned.
The patient's condition is
always changing.
6. Troubleshooting
Solution
• Take a measurement in a quiet
place.
• DO NOT talk while taking a
measurement.
• DO NOT apply this device on a
crying patient.
Contact your OMRON retail outlet
or distributor.
Replacing all batteries with 2 new
batteries is recommended. Refer
to sub-section 2.1.
• Immediately replace all
batteries with 2 new batteries.
Refer to sub-section 2.1.
• Confirm the batteries are properly
inserted with polarities correctly
aligned. Refer to sub-section 2.1.
NEVER diagnose or treat a patient
based only on the result. Consult with
a physician if you are concerned.
EN
EN18