Ameda Penguin Instructions For Use Manual

Ameda Penguin Instructions For Use Manual

Deluxe nutritional warmers
Hide thumbs Also See for Penguin:

Advertisement

Ameda Penguin
Deluxe Nutritional Warmers
Instructions for Use
1-15
EN
Instrucciones de uso
17-32
ES
Mode d'emploi
33-48
FR
Penguin
®
Nutritional Warmer

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Penguin and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Ameda Penguin

  • Page 1 Ameda Penguin Deluxe Nutritional Warmers Instructions for Use 1-15 Instrucciones de uso 17-32 Mode d’emploi 33-48 Penguin ® Nutritional Warmer...
  • Page 2 WELCOME TO THE AMEDA FAMILY We are excited to help you meet your nutritional warming needs. With the Ameda Deluxe Single- Well or Four-Well Penguin® Nutritional Warmers, you can safely warm feedings and protect vital nutrients. The Ameda Deluxe Nutritional Warmers work with the patented design Ameda Therma-Liner®...
  • Page 3: Table Of Contents

    8 Maintenance ................................12 9 Servicing And Spare Parts ............................12 10 Warranty .................................12 11 Technical Data, Storage And Operating ....................... 13 12 Conditions ................................13 13 Ameda Accessories .............................. 13 Warning Temperature Limitation For Transport/Storage Serial Number Caution Humidity Limitation For Transport/Storage...
  • Page 4: Warnings And Cautions

    1. WARNINGS AND CAUTIONS IMPORTANT SAFEGUARDS When using electrical products, the following basic safety precautions should always be followed. READ ALL INSTRUCTIONS BEFORE USING THIS PRODUCT DANGER To reduce the risk of electrocution: • DO NOT place or store where warmer can fall or be pulled into a sink. •...
  • Page 5 WARNING To avoid fire, electrocution, serious burns or injury. • ALWAYS operate warmer on a flat, horizontal surface. • Close supervision is necessary if product is used near children or infants. • Use product for its intended use ONLY, as described in this manual. •...
  • Page 6: Intended Use

    3. PRODUCT DESCRIPTION Penguin Nutritional Warmer Your Penguin Nutritional Warmer is a feeding warming device and is customized with settings and warming profiles according to your unique facility needs. The nutritional warmer operates with gentle vibration which helps preserve and suspend feeding nutrients and promote even warming.
  • Page 7: Set-Up Instructions

    • -abc/123 • Enter button=advance to next entry • On/Off button=edits complete, return to Standby mode 4.3 MODES STANDBY MODE The Penguin Nutritional Warmer will be in Standby mode and cycle through the following screens whenever powered on and not warming:...
  • Page 8 WARMING MODES Fridge to Feed Feed Temp range of 90-100°F/32-38°C Average Time 5-12 minutes depending on volume Frozen to Feed Feed Temp range of 90-100°F/32-38°C Average Time 10-18 minutes depending on volume Thaw After Thaw Temp range of 38-45°F/3-7°C Time 15-25 minutes depending on volume READY TO FEED MODE - HOLD TIME (Fridge to Feed and Frozen to Feed only) The feeding is ready when the screen background turns green with the following words displayed: REMAINING HOLD TIME...
  • Page 9: Therma-Liner Preparation

    DISPLAYED ERRORS If the Penguin Nutritional Warmer encounters an error, the error will be displayed on a red screen with a warning tone. If this should occur, discontinue use of the warmer and call Ameda CustomerCare at 1.877.99.AMEDA (26332).
  • Page 10: Inspecting Before Operation

    8. Please remove the feeding from the Therma-Liner and set the Therma-Liner aside until next use during the shift. 9. Press On/Off to return the Penguin to Standby mode. * Note that if a wrong entry was made, you may press the...
  • Page 11: Thaw Frozen Milk To Cool Liquid State

    6.2 THAW FROZEN MILK TO COOL LIQUID STATE Caution: the Penguin Nutritional Warmer is not recommended for thawing less than 15 mL of breast milk in storage bags! 1. Scroll to choose the container type and size of interest. 2. Press Enter.
  • Page 12: Maintenance

    Ameda’s sole obligation under this limited warranty shall be to repair or replace, at Ameda’s option, any Product that is determined to be defective by Ameda and determined to be covered by this express limited warranty.
  • Page 13: Technical Data, Storage And Operating

    11. TECHNICAL DATA, STORAGE AND OPERATING CONDITIONS Penguin Single Well Nutritional Warmer Power Rating: 220-240VAC, 50/60 Hz, 1.25A Relative Humidity: 80% up to 31°C to 50% to 40°C / 80% up to 87.8°F to 50% up to 104°F Altitude: 2000 meters Operating Temperature: 5 to 40°C/41 to 104°F...
  • Page 14 EQUIPMENT. Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Penguin Nutritional Warmer, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
  • Page 15 At 80 MHz and 800 MHz, the higher frequency electromagnetic site survey should be considered. If the measured range applies. field strength in the location in which the Ameda Penguin Warmer is NOTE 2 These guidelines may not apply in all situations. Electro-...
  • Page 16 Distribución en otros países: para obtener información o comentarios sobre el producto, llame a su distribuidor local o al lugar donde compró el producto. Para obtener una lista de los distribuidores en su país, visite www.ameda.com. Transport/Storage Operation Transporte/almacenamiento...