1.YOUR NEW NEOPIPETTA The neoPipetta has been carefully designed and manufactured as a lightweight, cordless Pipette Filler. neoPipetta offers simple, efficient pipetting performance and maximum pipetting comfort. It can be used with all plastic or glass pipettes between 0.1-100ml. 2. Charging the neoPipetta Before the first use the battery must be charged for 10 hours.
4. NEOPIPETTA DESCRIPTION 4.1. Operating buttons The neoPipetta has two operating buttons located at the front of the handle. The upper button enables aspiration for as long as the button is held down. The lower button allows for dispensing. The aspiration and dispensing speeds can be fine-tuned during operation by applying varying finger pressure on the operating buttons.
5.3. Aspiration Select a speed and fill the pipet by pressing the aspirating button. Fill the pipet just above the desired level, check level by the meniscus. Remove the serological pipet from the sample. Adjust the volume of the sample by pressing the dispensing button until the meniscus is at the required level.
Fig.2 Product D-6026 Silicone adapter D-6040 Nose cone for neoPipetta D-6028 Filter 0.45µm D-6029 Battery The neoPipetta cannot be operated without a filter. Wetting of filter due to overfilling will slow or prevent fluid movement. If overfilling occurs, change the filter immediately. Only use filters supplied by the manufacturer. 6.3.
8. TROUBLE SHOOTING Problem Cause Solution Reduced aspiration capacity - Filter blocked, dirty or damaged -Replace the filter or the neoPipetta is blocked -Battery low -Recharge Serological pipet drips -Serological pipet not securely attached -Reattach serological pipet -Silicone adapter or nose cone loose -Reassemble silicone adapter or nose cone -Silicone adapter or nose cone damaged...
Need help?
Do you have a question about the neoPipetta D-6019 and is the answer not in the manual?
Questions and answers