HP SCR-1P Series Operating And Service Manual page 39

Hide thumbs Also See for SCR-1P Series:
Table of Contents

Advertisement

5-16.
LOAD REGULATION
5-17
.
CONSTANT VOLTAGE. Proceed as
follows:
a. Connect the 48 -ohm load resistor across the output terminals
and
the
differential voltmeter across the -S
and +S tenninals.
b. Tum front-panel CURRENT controls fully clockwise (maximum).
c. Tum the front-panel
VOLTAGE controls until
rront-panel
ammeter
indi-
cates2. 5 amperes.
d
.
Record voltage indicated
on
differential voltmeter.
e. Disconnect load
resistor.
f. Differential voltmeter indication
should change
by less
than-
120
mvdc.
5
-
18.
CONSTANT CURRENT. Proceed
as follows:
a.
Connect test setup shown in figure 5-3.
b. Turn front-panel VOLTAGE controls fully clockwise (maximum)
.
c. Turn front-panel CURRENT controls
until
front-panel ammeter indicates
2. 5 amperes.
d
.
Record
voltage indicated
on
differential voltmeter.
e. Close
the
shorting switch.
f. Differential voltmeter indication
should change by
less than
25
mvdc.
5-19
.
RIPPLE AND NOISE
5-20
.
Proceed as follows:
a
.
Connect the
48 -ohm load resistor across the output terminals
and
the
ac voltmeter across the
-S
and
+S terminals.
b. Tum front-panel
CURRENT
controls fully
clockwise
(maximum)
.
c. Connect
the variable voltage
transformer between
the input power
source
and
the power
supply
power input.
Adjust the
variable voltage transformer
to
125
vac.
d
.
Tum front-panel
VOLTAGE controls until front-panel
ammeter
indicates
~.
5amperes.
e. The ac voltmeter should indicate
less
than 240 mvrms
.
5-4