Installing the G1330B Thermostat
Site Requirements
on the power cords which connect to the rear of the two instrument is
identical for all types of power cord.
Never operate your instrumentation from a power outlet that has no
ground connection. Never use a power cord other than the power cord
designed for your region.

Bench Space

The thermostatted autosampler dimensions and weight (see Table 1) allow
the instrument to be placed on almost any laboratory bench. The instrument
requires an additional 25 cm (10 inches) of space on either side for the
circulation of air, and approximately 8 cm (3.1 inches) at the rear for
electrical connections. Ensure the thermostatted autosampler is installed in a
level position.
If a complete Agilent 1100 Series system is to be installed on the bench, make
sure that the bench is designed to carry the weight of all the modules. For a
complete system including the thermostatted autosampler it is recommended
to position the modules in two stacks, see "Optimizing the Stack
Configuration" on page 25. Make sure that in this configuration there is 25 cm
(10 inches) space on either side of the thermostatted autosampler for the
circulation of air.

Environment

Your thermostatted autosampler modules will work at ambient temperatures
and relative humidity as described in Table 1.
Do not store, ship or use your thermostatted autosampler under conditions
where temperature fluctuations may cause condensation within the
thermostatted autosampler electronics. Condensation will damage the system
electronics. If your thermostatted autosampler was shipped in cold weather,
leave it in its box, and allow it to warm up slowly to room temperature to avoid
condensation.