PENLON ESO2:
QUICK START GUIDE
CONNECT AND SET UP THE DEVICE
1
How do I connect the ventilator?
(Image A)
• Connect medical gas hoses to oxygen
and air and connect to mains power
• Connect breathing system to
• Switch
on, indicator turns green
SET THE CONTROLS
3
Volume Controlled Ventilation
(Image A & C) START UP MODE
• Turn device On
• Set Oxygen to 9L/min
• Set Air to 3L/min
• Attach patient end of breathing system to
test lung / bag observe bellows fill
• Select 'ADULT DEFAULTS' on the display
screen
• Select 'VOLUME' on the display screen
and press once
• Set Volume to 400 mls, Rate to 20,
I:E = 1:2 and PEEP to 15
• Press Volume twice to start ventilation
• Check Oxygen reads 80% and High
pressure alarm at 35 cmH
O
2
• Only then attach to patient
This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be used in conjunction with the manufacturer's Instructions for Use
This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be used in conjunction with the manufacturer's Instructions for Use
A
1
and
2 3
Display screen
Oxygen set to 9
Air set to 3
Filter
8
7
4
5
6
Filter
On switch
Bellows
Dial (turn to select, push to confirm)
TYPE B ANAESTHETIC MACHINE TYPE
This machine does not use an absorber
2
How do I set up the breathing
system and the HME filter?
(Image B)
• Attach catheter mount to the HME filter
and then to the Y piece of the breathing
circuit
• Ensure that the other end of the system is
A
A
attached to
and
B
shown
B
Catheter Mount
HME Filter
C
ADULT DEFAULT screen view
with filters as