Page 1
Quick Reference Guide P/N 3300995, Rev. C April 2003 Model 3500 Transmitter (9-wire) or Model 3300 Peripheral Installation Instructions for Rack-Mount For online technical support, use the EXPERT ™ system at www.expert2.com. To speak to a customer service representative, call the support center nearest you: •...
Page 3
Micro Motion support center. Installation kit For mounting in a rack, the Model 3300/3500 installation kit includes the following parts: • One DIN standard 41612, Type D connector for input/output wiring, with solder tails or screw terminals •...
Page 4
1 U (1 HE) 3 U (3 HE) 1 U (1 HE) 3 U (3 HE) Dimensions The Model 3300/3500 has the following dimensions (see Figure 2, page 4): • Height: 128 mm (3 U or 3 HE) • Width: 142 mm (28 HP or 28 TE) •...
Page 5
8 33/64 (216,2) Flowmeter cable lengths Maximum cable length from the sensor to the Model 3500 transmitter is 1000 feet (300 meters). If you are installing the Model 3300 applications peripheral in combination with a transmitter, the maximum cable length from the transmitter’s frequency output to the Model 3300’s frequency input is...
Page 6
2. Install the input/output wiring connector at 4 HP (4 TE) from the neighboring unit or from the edge of the rack. 3. (Model 3500 only) Install the keyed sensor wiring connector at 16 HP (16 TE) from the neighboring unit or from the edge of the rack.
Page 7
Installing the Model 3300/3500 in the rack 1. Align the Model 3300/3500 with the guide rails. 2. Slide the Model 3300/3500 into the rack. Make sure the pins on the back panel make contact with the wiring connectors. 3. Tighten the supplied captive screws to secure the front panel of the...
Page 8
Input/output wiring terminal card STEP 5. Connecting the Model 3500 to the sensor To connect the Model 3500 transmitter to a Micro Motion sensor, follow the steps below. If you are installing the Model 3300 applications peripheral, this step is not required.
Page 9
2. Prepare the cable according to the instructions in Micro Motion’s 9-Wire Flowmeter Cable Preparation and Installation Guide. 3. Ensure that the cable has 360° shielding, continuous from the transmitter to the sensor’s junction box. Two methods can be used: •...
Make sure power supply voltage matches voltage that is indicated on power supply wiring terminals. See Figure 6, page 10. Referring to Figure 6, page 10, connect the Model 3300/3500 to a power supply as follows: 1. Use 18 to 14 AWG (0,75 to 2,5 mm ) wire.
Page 11
Figure 6. Power supply wiring terminals Power supply wiring terminals Terminal designations for DC power Terminal designations for AC power...