AngioDynamics C3 WAVE Quick Reference Manual page 4

Hide thumbs Also See for C3 WAVE:

Advertisement

Warning Screen Pop-Ups
WARNING
PICC Lead Off
ECG Lead Off
Internal System Error
Charger Error
Internal Power Error
Hub Disconnected
Q U I C K R E F E R E N C E G U I D E
CONDITION
PICC lead disconnected from the patient.
App in "internal mode".
ECG lead disconnected from the patient.
I2C bus failure
Battery charger failure
3.3V power rail goes below 3.1V
Wireless connection with the hub is lost
PERSISTENCE
Skippable -
cleared by pressing "Dismiss"
Skippable -
cleared by pressing "Dismiss"
Unskippable -
cleared by power cycling the device
Unskippable -
cleared by removing and re-inserting
the DC power connector
Unskippable -
cleared by power cycling the device
Unskippable -
cleared by reconnecting the hub
SCREENSHOT
C3 WAVE