3. Clean all parts using a soft cloth with cleaning solution.
4. Thoroughly rinse all parts under running water.
5. Prepare a disinfectant bath consisting of water and a disinfectant.
6. Put all parts to be disinfected in the disinfectant bath.
7. Thoroughly rinse all parts under running water.
8. Dry all parts in the air or in a drying cabinet (temperature: max.
60 °C). Do not expose to direct sunlight. Do not place the mask/
helmet adapter down on the face seal.
9. Grease the guides on the mask/helmet adapter with Voltalef or
Molykote 111.
10. Mount the accessories, if necessary.
NOTICE
i
i
The full face mask can also be cleaned and disinfected in a
washing machine. Further information can be obtained from
Dräger.
4.3
Checks
Check the full face mask according to the inspection and maintenance
intervals.
4.3.1
Visual inspection of the speech diaphragm
The speech diaphragm and O-ring must be clean and undamaged.
Otherwise clean or replace (see Chapter 4.4.4 on Page 19).
4.3.2
Visual inspection of the inhalation valve
The inhalation valve can be seen in the connector through the
visor. Check if both blades of the valve disc move freely by blowing
into the connector. Replace the valve disc, if required (see
Chapter 4.4.6 on Page 19).
Panorama Nova S - P
4.3.3
Visual inspection of the exhalation valve
1. Remove the exhalation valve cover
from the connector.
2. Press in one side and remove the
spring carrier.
3. Pull the valve disc out of the guide.
4. The valve disc and valve seat must be
clean and undamaged. Otherwise
clean and replace.
5. Moisten the valve disc with water and insert it. The valve disc must
slide into the guide of its own accord.
6. Set the spring carrier in a way that both sides snap into place on the
sides. It is marked with "L" = left and "R" = right and sits at an angle.
4.3.4
Leak tests
Perform the leak test with a suitable test unit (e.g. from the Testor or
Quaestor series).
(1) Checking leak tightness
1. Moisten the inside of the face seal with water and mount the full
face mask on the test head with mask holder S.
2. If necessary, inflate the test head until the face seal fits
everywhere.
3. Seal the connector with the adapter.
4. Moisten the exhalation valve if necessary.
5. Generate a negative pressure of 10 mbar.
The full face mask is considered leak-tight if the change in pressure
after 1 minute does not exceed 1 mbar. Otherwise proceed with (2).
(2) Check with sealed exhalation valve
1. Remove the exhalation valve disc.
2. Insert the exhalation valve plug and fasten it with the spring carrier.
3. Generate a negative pressure of 10 mbar.
4. Remove the exhalation valve plug if the pressure difference is less
than 1 mbar after 1 minute, insert a new valve disc and repeat the
leak test. Otherwise, continue with (3).
(approved by Dekra Exam)
Maintenance
L
R