Calibrating Redox Transmitters And Checking Connected Components - JUMO 202711/10 Operating Manual

Simulator for ph/redox
Table of Contents

Advertisement

4.5

Calibrating redox transmitters and checking connected components

In principle, the process used for checking a redox transmitter is similar to the process used for checking
a pH transmitter – as described on page 10.
Depending on the measuring span of the redox transmitter to be calibrated, set values between 7 pH
and 0 pH on the simulator's pH selection switch. Please refer to the redox potential table below for se-
lecting the correct mV value.
Example:
The measuring span of the redox transmitter is 0 to 100 mV.
Zero point adjustment in the case of switch position 7 pH = 0.00 mV 2 mV.
Redox potential table
Switch position pH
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Potential ±1 %
+414 mV
+355 mV
+296 mV
+237 mV
+177 mV
+118 mV
+59 mV
0 mV
-59 mV
-118 mV
-177 mV
-237 mV
-296 mV
-355 mV
-414 mV