4000 SERIES AUTOMATIC SWITCHOVER Allied Healthcare Products, Inc. MANIFOLD 1720 Sublette Avenue - St. Louis, MO 63110 Operation & Maintenance Manual Telephone: 800-444-3954 - Fax: 314-771-1806 Catalog No. S188-354-001 Rev. B MANIFOLDS ARE AVAILABLE FOR THE FOLLOWING GASES: OXYGEN CARBON DIOXIDE...
Table of Contents Subject Page Introduction Key Elements Manifold Operation Factory Settings Schematic Diagram Flow Charts Site Selection Installation Manifold Outline Method Of Operation Safety Features Remote Alarm Wiring Diagram Leak Test Ordering Information Repair Parts List Trouble Shooting 17-18 Limited Warranty Back Cover...
Page 3
WARNING ALL PRESSURE ADJUSTMENTS SHALL BE MADE BY AUTHORIZED PERSONNEL ONLY. INCORRECT PRESSURES CAN DISRUPT GAS DELIVERY TO LINE. WARNING THE MANIFOLD SYSTEM MUST BE LOCATED IN A WELL VENTILATED AREA. NITROUS OXIDE CAN CAUSE UNCONSCIOUSNESS AND DEATH. OXYGEN AND N2O AND CAN GREATLY INTENSIFY A SPARK OR FIRE.
Manifolds may be obtained in these gas services as standard units. The manifold unit and header bar sections can be ordered complete by calling your Allied Sales Representative. Header bar sections can also be obtained by ordering the number and type needed to complete the manifold from the list on Page 15. NFPA requires at least the number of cylinders for an average twenty-four (24) hour day’s supply per bank, with a minimum of two cylinders per bank.
Manifold Key Elements SERVICE LINE PRESSURE LINE RELIEF OUTLET EXHAUST VENT LINE PRESSURE GAUGE CHECK CHECK VALVES(NOT VALVES(NOT SHOWN) SHOWN) LEFT LINE PRESSURE REGULATOR RIGHT LINE PRESSURE REGULATOR BALL VALVE PC BOARD SOLENOID VALVE LT. SOLENOID BANK VALVE RT. BANK POWER SUPPLY PNEUMATIC...
Key Elements 1. A pressure control manifold that maintains constant pressure to the pipeline and enables smooth uninterrupted switchover from the “in use” to the “reserve” cylinders. 2. A 115VAC 0.2 Amps / 230 VAC 0.1 Amps input, 5/15 VDC output power supply. Power supply includes dry contacts for local and remote alarm connections.
WARNING ALL PRESSURE ADJUSTMENTS SHALL BE MADE BY AUTHORIZED PERSONNEL ONLY. INCORRECT PRESSURES CAN DISRUPT GAS DELIVERY TO LINE. Schematic Piping Diagram Schematic Piping Diagram...
90%. If the humidity is high and the temperature is low, frost and/or ice could interfere with the performance of the regulators. Allied Healthcare Products, Inc. (Allied) includes a set of heaters as standard equipment with carbon dioxide and nitrous oxide control units that will preheat gas from the cylinders minimizing the frost problem.
If damage is discovered during the uncrating and assembly process, contact the freight carrier and make a concealed damage report within seven (7) days. Allied will assist the claims process as much as possible, but it is the responsibility of the receiver to file any claims.
Manifold Outline Approximate floor space needed to install manifold and cross header bar sets. Cylinder Overall Width Cylinder Overall Width 1 X 1 17 inches 9 X 9 73.5 inches 2 X 2 22 inches 10 X 10 94 inches 3 X 3 22 inches 11 X 11...
Page 12
CAUTION DO NOT APPLY HEAT TO DELIVERY LINE UNION WHILE IT IS CONNECTED TO THE CONTROL UNIT. HEAT WILL HARM O-RINGS, CAUSING LEAKS AND MALFUNCTIONS. Check for proper venting to atmosphere as required by job specifications and/or NFPA 99, If the manifold is installed in an enclosed area, the safety relief valves must be vented to the outside atmosphere.
Adjust as needed and tighten. Cylinder Pigtails Allied offers two styles of flexible pigtails: coiled copper (oxygen and nitrous oxide) and braided stainless steel (nitrous oxide and all other gases). Refer to the Optional Equipment Section or contact your Allied Sales Representative. The braided pigtail offers greater ease of handling the gas cylinders.
“Reserve” to “in Service”. Alarm Systems Allied Healthcare Products, carries a complete line of alarm systems which can be used in conjunction with the 4000 Series Manifold, to provide the required visual and audible signals, when bank switching occurs. For remote alarm electrical connection, use Electrical Wiring Diagram shown on page 13. The 4000 Manifold System, when used in conjunction with these alarm systems, complies with CSA Z305.1-M1984.
Page 15
Function Identification The bank LED’s and pressure gauges on the cabinet face are clearly marked to explain their function. The 4000 Series Manifold header bars are designed to assure that only cylinders containing the proper gas can be connected to them. All header bar connections and pigtails comply with CGA Standards. CAUTION DO NOT ATTEMPT TO RETROFIT THE CONTROL UNIT TO ACCOMMODATE OTHER GASES THAT THE CONTROL UNIT WAS NOT DESIGNED FOR.
Wiring Diagram TO LEFT BANK TO RIGHT BANK SOLENOID VALVE SOLENOID VALVE POWER SUPPLY DC POWER FROM POWER SUPPLY GREEN-GROUND WHITE-NEUTRAL GREEN LEDS BLACK-LINE MANIFOLD PC BOARD YELLOW TERMINAL LEDS BLOCK 115V AC-0.2A 230V AC-0.1A LEDS OUTPUTS 50/60 HZ AC POWER CONNECTION 2 #18 GA TO MASTER ALARM PANEL "RESERVE IN USE"...
Leak Test In order to determine whether any leaks exist between header bar sections or at the central pipe system connection, the system must be pressurized. Either oil-free dry air or oil-free dry nitrogen may be used for new installations. In the case of either a medical air or nitrogen 4000 Series Manifold, the actual service gas may be used to perform the leak tests.
Ordering Information The 4000 Series Manifold has two (2) outlet (line) regulators and one (3) safety relief vent. The 4000 Series Manifold will shift banks automatically and will automatically reset when the depleted cylinders are replaced. MAIFOLDS: Catalog Numbers System 86-71-2100 Control Unit Only Oxygen...
Repair Parts List Catalog No. Description Qty. Per Kit 86-90-3041 High Pressure Relief Valve 86-90-3042 Line Pressure Relief Valve Nitrous Oxide, Carbon Dioxide, Medical Air & Oxygen 86-90-3043 Line Pressure Relief Valve Nitrogen 86-90-3001 High Pressure Hand Valve Oxygen 86-90-3002 High Pressure Hand Valve Nitrous Oxide 86-90-3003...
Only some minor checks and repairs are recommended to be made in the field. Components removed from the manifold system for maintenance must be serviced, repaired and tested by properly qualified medical service personnel only. Original manufacturer’s parts, as supplied by Allied, must be used in the maintenance of manifolds.
Page 21
Trouble Probable Cause Remedy or Check Loose Regulator cap Tighten cap Leaking High pressure Regulator Regulator Diaphragm Leaks Replace Diaphragm Loose Regulator cap Tighten cap Leaking Low pressure Regulator Regulator Diaphragm leaks Replace Diaphragm Filter upstream of high pressure Remove inlet valve, inspect and clean Low Gas Flow regulator blocked filter.
Allied within warranty period of product purchased, that the product is found by Allied to have been defective at the time of such shipment, that the product has been installed and/or operated in accordance with Allied’s instructions, that no repairs, alterations or replacements have been made by others without Allied’s written approval, and that...