Bartels Mikrotechnik microComponents mp6 Series Operating Manual

Bartels Mikrotechnik microComponents mp6 Series Operating Manual

Micropump
Hide thumbs Also See for microComponents mp6 Series:

Advertisement

1
Operating Manual for
Micropump mp6-series
Rev. 3.1 (08.2020)

Advertisement

Table of Contents
loading

Summary of Contents for Bartels Mikrotechnik microComponents mp6 Series

  • Page 1 Operating Manual for Micropump mp6-series Rev. 3.1 (08.2020)
  • Page 2: Table Of Contents

    Content General ..........................................3 Declaration of conformity ..................................3 Description of functions ..................................3 Proper use ........................................4 Intended purpose ..................................... 4 Misuse......................................... 4 Staff selection and qualification ................................ 4 About this operating manual ................................4 Technical specifications ..................................... 5 Technical Data of the mp6-hyb ................................
  • Page 3: General

    Do not open the housing of the micropump and the controllers. In those cases, Bartels Mikrotechnik cannot issue a guaranty anymore. Please keep this manual safe and give a copy to all users.
  • Page 4: Proper Use

    Do not make any modifications or extensions to the pump or controller without the prior written consent of the manufacturer. Such modifications may impair the safety of the unit and are prohibited! Bartels Mikrotechnik GmbH rejects any responsibility for damage to the unit caused by unauthorized modifications to the pump and risk and liability are automatically transferred to the operator.
  • Page 5: Technical Specifications

    Technical specifications Technical Data of the mp6-hyb mp6-hyb Order code: mp6-hyb Pump type piezoelectric diaphragm pump Number of actuators 30 x 15 x 3,8 mm Dimensions without connectors 1.1811 x 0.5906 x 0.1498 in. Weight barbed tube clip, Fluidic connectors (outer diameter 1.9 mm, length 3.5 mm) Electric connector flex connector 1.25 mm pitch...
  • Page 6: Technical Data Of The Mp6-Liq

    Technical Data of the mp6-liq 1, 2 mp6-liq Order code: mp6-liq Pump type piezoelectric diaphragm pump Number of actuators 30 x 15 x 3,8 mm Dimensions without connectors 1.1811 x 0.5906 x 0.1498 in. Weight barbed tube clip, Fluidic connectors (outer diameter 1.9 mm, length 3.5 mm) Electric connector flex connector 1.25 mm pitch...
  • Page 7: Technical Specifications Mp6-Pi

    Technical specifications mp6-pi mp6-pi Order code: mp6-pi Pump type piezoelectric diaphragm pump Number of actuators 30 x 15 x 3,8 mm Dimensions without connectors 1.1811 x 0.5906 x 0.1498 in. Weight barbed tube clip, Fluidic connectors (outer diameter 1.9 mm, length 3.5 mm) Electric connector flex connector 1.25 mm pitch Power consumption...
  • Page 8 mp-x, mp6-EVA, mp6-OEM, mp6-QuadOEM and mp6- Suitable pump driver QuadKEY Typical values of flow and back pressure for selected media (values measured with mp-x: 100 Hz, 250 V, SRS): Liquids – water 8 µl/min – 8000 µl/min Controllable flow range ��...
  • Page 9: Technical Specifications Mp6-Pp

    Technical specifications mp6-pp mp6-pp Order code: mp6-pp Pump type piezoelectric diaphragm pump Number of actuators 30 x 15 x 3,8 mm Dimensions without connectors 1.1811 x 0.5906 x 0.1498 in. Weight barbed tube clip, (outer diameter MIN 1.77 mm Fluidic connectors - MAX 1.85 mm, length 3.5 mm) Electric connector flex connector 1.25 mm pitch...
  • Page 10: Technical Specifications Mp6-Gas

    Technical specifications mp6-gas mp6-gas Order code: mp6-gas Pump type piezoelectric diaphragm pump Number of actuators 30 x 15 x 3,8 mm Dimensions without connectors 1.1811 x 0.5906 x 0.1498 in. Weight barbed tube clip, Fluidic connectors (outer diameter 1.9 mm, length 3.5 mm) Electric connector flex connector 1.25 mm pitch Power consumption...
  • Page 11 Typical values of flow and back pressure for selected media (values measured with mp-x: 300 Hz, 250 V, SRS): Gases 20 ml/min (300 Hz) typ. volume flow ��̇ (p=0) 5, 8 typ. back pressure p (��̇ = 0) 100 mbar (300 Hz) Typical values.
  • Page 12: Final Inspection

    Final inspection After production, the micropumps have to pass a final inspection. They are tested concerning the maximum flow and back pressure. Measurement conditions: Pumping media: distilled water Temperature: room temperature 23°C Controller: extended box mp-x Electrical Input: amplitude of 250 Vpp and SRS-Signal with 100 Hz (for liquids) and 300 Hz (for gases) Measurements with sensors: range: 0 - 10 ml/min...
  • Page 13: Connecting The Mp6-Series To The Cable

    Bartels Mikrotechnik. Because of the mechanical fixation inside the electrical connector and due to the use of elevated driving voltages it is ensured that the pump performance is not affected by the surface corrosion. In a long term, we will work on improving the production process. Until then, referring to the unrestricted pump performance we ask our customers to excuse visible defects on the electrical connector.
  • Page 14: Cleaning The System

    1 Änderungen vorbehalten If the pump should be removed again, the Molex connector needs to be opened before removal! In order to prevent damage to the flexible cable, the following points should be considered for the final pump assembly: the flexible connector must not be bent around sharp edges or kinked the flexible connector must not be bend on top or bottom of the pump the Molex connector is not water tight, additional sealing with e.g.
  • Page 15: Typical Operation Parameters

    Typical operation parameters mp6-hyb and mp6-pi The description here applies to the medium water. For the delivery of gases, please read the section to mp6-gas. Please note that the regular mp6-hyb, as well as the mp6-liq can also pump gases, but has not been specially measured for this purpose.
  • Page 16: Combination Of Micropumps

    If lower flow rates are required, it is advisable to use a restrictor for throttling and to operate the pump at the recommended parameters. As a general experimental approach, we recommend a high amplitude at varying frequencies. The performance depends on the environmental conditions. Combination of micropumps The micropumps can be combined to achieve higher flow rates or pressure than the single unit can generate.
  • Page 17 Visit our Website Contact Data: www.bartels-mikrotechnik.de Bartels Mikrotechnik GmbH for further information on applications. Konrad-Adenauer-Allee 11 Tutorials and helpful answers to frequently asked 44263 Dortmund Germany questions can be found in our FAQ www.bartels-mikrotechnik.de http://blog.bartels-mikrotechnik.de info@bartels-mikrotechnik.de or on our YouTube channel Tel: +49-231-47730-500 https://www.youtube.com/user/BartelsMikrotechnik...

Table of Contents