Checking The Sensitivity Of Counters - Fluke PM 6666 Operator's Manual

Timer / counter
Table of Contents

Advertisement

Page: 44
Low Impedance Inputs (50Ω)
If You Have a Calibrated Signal Source
Adjust the signal-source to the sensitivity limit of the
counter.
Connect it directly to the input of the counter.
Check that the counter is operating correctly.
If You Don't Have a Calibrated Signal Source
Use either of the following methods
Method 1
Connect the output of the signal-source directly to the in-
put of the counter.
Turn off AUTO and Set the counter to maximum sensitiv-
ity (if adjustable).
Adjust the amplitude of the signal-source to the minimum
level accepted by the counter.
PM 6666 - OPERATORS MANUAL
APPENDIX 1
Disconnect the cable from the counter and connect it to
the oscilloscope.
Read the amplitude on the oscilloscope.
Check that the reading is the same as, or less than, the
sensitivity level in the counter specifications.
Method 2
Connect the signal-source to the oscilloscope.
Adjust the output amplitude of the signal-source until the
oscilloscope indicates the sensitivity limit in the counter
specifications.
Disconnect the cable from the oscilloscope and connect
it to the counter.
Turn off AUTO and set the counter to maximum sensitiv-
ity (if adjustable).
Check that the counter is operating correctly.
These procedures ensure unambiguous measurements of
the signal voltage at the input of the counter.