8 Sensor Maintenance and Precaution 8.1 Methods to Cleaning the ST-565 Series Probe ......8.2 Storage ..........
Page 4
Pyxis Lab, Inc. Standard Limited Warranty Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new).
1 Introduction The ST-565 Series inline fluorometers are patented to Pyxis Lab Inc. These devices utilized UV light fluores- cence to detect Halogenated Tolytriazole (HST with ST-565) and Tolytriazole (TTA with ST-565T) content in water. The results are instantaneous and, with proper application understanding, results are highly reliable.
Enclosure Rating IP67 Regulation With Pyxis’s continuous improvement policy, these specifications are subject to change without notice. 3 Unpacking Instrument Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipping. Verify that all accessory items are included. If any item is missing or damaged, please contact Pyxis Lab Customer Service at service@pyxis-lab.com.
4 Installation 4.1 Piping The ST-565 Series probe can be installed in an inline flow application using the included Pyxis ST-001 Inline Tee Assembly. The ST-001 offers 3/4” FNPT thread or socket weld adapters with unions for easy installation and sensor maintenance. To properly install the ST-565 Series probe into the ST-001 Tee Assembly, follow the steps below: 1.
4.3 Connecting via Bluetooth A Bluetooth adapter (P/N: MA-WB) can be used to connect a ST-565 Series probe to a smart phone with the uPyxis® Mobile App or a computer with the uPyxis® Desktop App. Figure 3. Bluetooth connection to ST-565 Series probe 4.4 Connecting via USB...
5.2 Connecting to uPyxis® Mobile App Turn on Bluetooth on your mobile phone (Do not pair the phone Bluetooth to the ST-565 Series probe). Open uPyxis® Mobile App. Once the app is open the app will start to search for the sensor. Once the uPyxis®...
When connected, the uPyxis® Mobile App will default to the Calibration screen. From the Calibration screen, you can perform calibrations by pressing on Zero Calibration, Slope Calibration, and 4–20mA Span. Follow the screen instructions for each calibration step. Figure 7. ST-565 Series User Manual service@pyxis-lab.com | +1 (866) 203-8397...
Clean message will be shown. If the probe is partially fouled, a yellow Becoming Dirty message will be shown. If the probe is severely fouled, a red Dirty message will be shown. In this case, follow the procedure in the Methods to Cleaning the ST-565 Series Probe section of this manual. Figure 8.
Desktop application. Double click the uPyxis.Setup.exe file to install. Figure 10. Click Install to start the installation process. Follow the screen instructions to complete the USB driver and uPyxis® installation. ST-565 Series User Manual service@pyxis-lab.com | +1 (866) 203-8397...
Desktop App will default to the Information screen. On the Information screen you can set the information description for Device Name, Product Name, and Modbus Address, then click Apply Settings to save. Figure 12. ST-565 Series User Manual service@pyxis-lab.com | +1 (866) 203-8397...
After the device has been calibrated and installation has been completed, to check diagnosis, click on Diag- nosis. When in the Diagnosis screen you can view the Diagnosis Condition of the device. This feature may be used for technical support when communicating with service@pyxis-lab.com. Figure 14. ST-565 Series User Manual service@pyxis-lab.com...
– 20 mA = 10.0 ppm 7.2 Communication using Modbus RTU The ST-565 Series probe is configured as a Modbus slave device. In addition to the ppm Fluorescent Polymer and ppb PTSA values, many operational parameters, including warning and error messages, are available via a Modbus RTU connection.
Figure 15. Inline Probe Cleaning Solution Kit To clean the ST-565 Series probe, soak the lower half of the probe in 100 mL inline probe cleaning solution for 10 minutes. Rinse the ST-565 Series probe with distilled water and then check for the flashing blue light inside the ST-565 Series probe quartz tube.
10 Contact Us Pyxis Lab, Inc 1729 Majestic Dr. Suite 5 Lafayette, CO 80026 USA www.pyxis-lab.com Phone: +1 (866) 203-8397 Email: service@pyxis-lab.com ST-565 Series User Manual service@pyxis-lab.com | +1 (866) 203-8397...