6.1
The battery change symbol
Troubleshooting
• The batteries are too low or empty. Replace both batteries with new 1.5 V AAA LRO3 batteries.
Erratic measurements being displayed
• Cuff incorrectly fitted or not tight enough. Fit cuff correctly and tighten.
• You have changed position or the cuff has moved during the measurement. Keep still and do not
• You have an irregular heartbeat. Even if your heartbeat is only slightly irregular, repeat the
• You have spoken during the measurement, you are nervous or you are excited. Do not speak and
• You are not in the corrected seated position. Please assume the correct position when taking the
• A problem occurred during inflation or during the measurement. Repeat the measurement.
"EE" appears in the display
• The unit has not been inflated enough. Repeat the measurement
• A problem occurred during inflation or during the measurement. Repeat the measurement.
No display or display locked
• No batteries have been inserted, batteries wrongly inserted or batteries are completely empty.
• A malfunction or fault has occurred. Briefly remove the batteries and then replace them.
Please contact the service centre if the error message appears again when the measure-
ment is repeated.
move your arm or the cuff.
measurement. The unit is not suitable for people with an extremely irregular heartbeat.
breathe deeply two or three times in order to relax.
measurement.
Insert batteries, check battery position or replace with new batteries.