The manual manifold can be used as either main
supply or emergency reserve manifold (ERM). When
used as a emergency backup manifold the line
pressure regulator should be set slightly lower than
the primary supply pressure . This will prevent it
from feeding to the pipeline during normal operation
of the primary system. The line pressure regulator
can be increased to nominal distribution pressure in
the event of emergency.
Figure 1 - Manual Manifold General Arrangement
13
10
9 (Valve shown in
Closed Position)
5
4
2
8 (Valve shown in
Open Position)
12
11
7
3
6
1
14
Table 1: Main features/components
Item
Description
No
1
1 stage regulator
2
Cylinder content/contact gauge
3
Line regulator
4
Line gauge
5
Integral Non-return Valve Assembly
6
1st Stage Pressure Relief Valve
7
Line Pressure Relief Valve
8
Lockable Isolation Valve
9
Test point Isolation Valve
10
Pipeline connection point (22mm
OD Copper Tube)
11
Pressure Relief Exhaust Connection
point (28mm OD Copper Tube)
12
Termination Box For Remote Alarm
13
Pressure Switch
14
High pressure bank valves
15
Single Line contact Module
(Not shown)
16
Double Line contact Module
(Not shown)
17
Manifold Headers C/W NRVs
18
Tailpipes
Notes...
Item 11 is recommended for indoor installations.
Item 15 & 16 are used to connect alarm outputs to a
Medipoint medical alarms.