Preparation for Cleaning
Equipment and Materials
WARNING: Use only Invuity-approved equipment unless otherwise specified. DO NOT modify any system component
or accessory.
The following equipment is required to process devices per the instructions in this manual:
Warm water (See the Water Quality section)
•
Prepared, specially formulated cleaning agents (See the Cleaning Agents section)
•
Absorbent wipes
•
Soft, nonlinting cloth
•
Syringe
•
Non-abrasive, soft, flexible, synthetic bristle brushes
•
Washer-disinfector (complies with ISO 15883 series)
•
Cleaning agents as required by the washer-disinfector manufacturer
•
Instrument air
< 140 kPa [< 20 psi]
1
•
Oven
•
1
Instrument air is medical gas that falls under the general requirements for medical gases as defined by the National Fire Protection
Association (NFPA) Health Care Facilities Code, is not respired, is compliant with the American National Standards Institute (ANSI) /
International Society of Automation (ISA) Quality Standard for Instrument Air, and is filtered to 0.01 microns, free of liquids and
hydrocarbon vapors, and dry to a dew point of -40 °C [-40 °F].
IFU 14466 Rev A
3