Description
The Invuity Sterilization Tray is used during steam sterilization. The Sterilization Tray is available in plastic and metal
material and available in multiple sizes.
Indications for Use
The Invuity Sterilization Tray is intended to allow transport and sterilization of enclosed medical device(s).
Contraindications
There are no known contraindications.
Warnings and Precautions
Cleaning and Sterilization:
See "Warnings and Precautions" in the Decontamination and Sterilization section of this Instructions for Use.
•
Product Use:
Prior to use, carefully inspect the Sterilization Tray for functionality, visible cracks, loose or bent
•
components, or rough edges. Discard damaged devices according to hospital procedures and in accordance
with local, state and federal laws and regulations.
When loading instruments and accessories in the Sterilization Tray, adequately space instruments far enough
•
apart to avoid contact with other metal instruments or object which may dent, ding, scrape or otherwise
damage the surface of the instrument.
•
Refer to the Instructions for Use for the Invuity devices for additional information.
Safety:
•
Use of instruments for any purpose or in any manner other than described in this Instructions for Use may
cause instrument damage or failure, which could result in patient or user injury.
Directions for Use
1. Prior to use, carefully inspect the Sterilization Tray for functionality, visible cracks, loose or bent components, or
rough edges. Do not use damaged trays.
2. Carefully load Invuity Fiber Optic Cables, Retractors and/or Accessories in the Sterilization Tray.
3. Sterilize the Tray and contents according to the Decontamination and Sterilization section of this Instructions
for Use.
IFU 14466 Rev A
Sterilization Trays - Instructions for Use