Page 1
Moisture Analyzer MB65 MB66 MB67 User Manual (Printer is optional) AMTAST USA INC...
Page 2
I Instruction 1. Security Please reading the following before you use the Moisture analyzer a. The instrument can use for test the moisture of the sample, each improper operation may result in personal injury and equipment damage b. Please access the AC power accordance with the manual; c.
II. Installation 2.1 Standard Packing List Please check the packaging and accessories. Standard Quantity Note Moisture Analyzer Mainframe 1 set Wind Cover 1 piece The sample pan holder 1 piece Tray handle 1 piece Power cord 1 piece Aluminum sample tray 1 set 50 pcs of aluminum pan 100g standard weight...
Page 4
2.4 Adjust the level and connect the power There are two feet in front of the instrument, adjust it and keep the instrument level. Connect the instrument with power, put one side of power cord Please warm up more than 30minutes for keep it stable and accurate. III Operation 3.1.
Page 5
3.3. Keypad NAME Key Features ON/OFF Power on / off, Just off the internal DC power supply START / STOP Start heated/ stop heated TARE Weighing peeled, Settings exit MENU Enter the setting menu CONVERSION Switching measurement results ENTER Confirm the current setting, go to next Return to the weighing mode (short press), calibration RESET balance (long press)
Page 6
heated tube lights, fans turn on, heated symbol in the upper left flashing. While rolling on the screen displays the current temperature, has been heating time, moisture content of the current sample. 6. End of the test, heated lamp automatic shut off, the buzzer alarm. Then the sample is displayed on the moisture content, heating time, current temperature and the parameter setting mode.
Page 7
c. Gentle heating mode: ( ) This mode is suitable for low moisture content, surface casing is easy to form, easy to burn the sample. From the heat began to rise to the long time required to set temperature, heating up even slowly. Heating time user selectable.
Page 8
Then flashing “100.000”, put on the calibration weight, then it will display "= = = = = =", until it display “100.000” remove the weight, it will display "= = = = = = "again, Waiting it show" 0.000 ", the calibration is completed. IV Technical Parameters Model MB65 MB66 MB67 Specification 110g/5mg...
Page 9
Schedule: Assembly drawing Please installation the instrument according the picture Installation order from top to bottom :1 --2--3--4--5 1. Aluminum scale weighing tray 2. The triangle scale frame 3. Scale tray brackets 4. Wind cover 5. Stainless steel plate 3.8 the use of the printer and Settings (Optional) A.
Page 10
3.Each time the switch machine, please interval of 5 seconds; 4. Cover the heater, do not leave the weight on the pan, or halogen lamps will be crushed. Amtast USA Inc. 438 Skyline Dr E Lakeland FL.33809 U.S.A. E-mail: amtast1@gmail.com, info@amtast.com...