Lab Stand Calibrator (PNº 800873-3) ALWAYS check to make certain you have received all of the items listed above. If you have any questions or need assistance, contact your Sensidyne Distributor, or call (800) 451-9444 (727) 530-3602 Sensidyne Document No. F-PRO-1260 (Rev C)
This manual was prepared exclusively for the owner of the Gilian Lab & Field Calibrator System. The material within this manual is proprietary information of Sensidyne, LP and is to be used only to understand, operate, and service the instrument. By receiving this document, the recipient agrees that neither this document nor the infor-...
Operation Manual. Operation of a malfunctioning product, or a product requiring repair may result in serious personal injury or death. If repair is needed, contact Sensidyne Service to arrange for a Returned Material Authorization (RMA).
(i.e., the sampling media) downstream of the agnostic mode, allows complete pump performance flowtube which then gives direct pressure drop read- checkout. ings and flow information about the loading source. Sensidyne Document No. F-PRO-1260 (Rev C)
(V2) through flowtube and connector block (V1). If leakage is still present, it indicates there is leak- age in the system prior to (V2). Sensidyne Document No. F-PRO-1260 (Rev C)
3) Follow another imaginary line horizontally from the point on the plot until it intersects with the Y- axis. This is the actual flow rate corresponding to the observed flow rate. Sensidyne Document No. F-PRO-1260 (Rev C)
5) Switch the Load Selector Valve (V1) to the Low This completes the pump performance check. If the Flow Adjust position (V4). pump has passed, you may set the desired flow rate and proceed with sampling. Sensidyne Document No. F-PRO-1260 (Rev C)
Substitution of components may impair Intrinsic Safety. Use only with UL Listed Portable Air Sampling Pump Battery Pack. Gilian Product Sensidyne, Inc. Clearwater, FL Made in USA LFS-113 PUMP Figure 3.1 Typical Calibration Set-Up Sensidyne Document No. F-PRO-1260 (Rev C)
HFS-513) maintain a constant flow rate when switching be- • at 3 LPM for HFS-113 & GilAir (4 LPM for HFS- tween the “Bypass” and “High Flow Adjust” modes. 513) Sensidyne Document No. F-PRO-1260 (Rev C)
“Low Flow Adjust” reset to apply the proper pres- sure range. NOTE It is imperative that flow rates are in the approximate range, and flowtubes must be calibrated to determine this. Sensidyne Document No. F-PRO-1260 (Rev C)
“Fault” light is “On”, do not adjust the valve any further. The pump should stop within one (1) minute after the red light actuates. This test can be run in any flow range setting. Sensidyne Document No. F-PRO-1260 (Rev C)
“FM1” position. Se- cure with three screws (#15). 7) Install three coupling tubes (#16) from upper mani- fold to selector valve as shown, and one tubing (#18) from damper (#21) to selector valve. Sensidyne Document No. F-PRO-1260 (Rev C)
Leakage check the assembly before from the lower manifold. using. 6) Clean the flowtube, using lab grade alcohol in a sy- ringe through the bottom. 7) After cleaning, blow dry with compressed air or air dry. Sensidyne Document No. F-PRO-1260 (Rev C)
Sensidyne’s re- SERVICE OPTIONS The Sensidyne Service Department offers you a variety of service options that will help to increase your user confidence while minimizing costly interruptions and maintenance costs. These options include initial training, on-site technical assistance, and full factory repairs. Sensidyne has developed several programs which will allow you to select just the right options which are best suited to your applications and needs.
Page 28
1000 112th Circle N, Suite 100 • St. Petersburg, Florida 33716 (800) 451-9444 • (727) 530-3602 • (727) 539-0550 (FAX) • www.sensidyne.com Revision C • Document No. F-PRO-1260...