Correcting Sample Vial Delivery Problems - HP 7683 Installation Manual

Automatic liquid sampler
Table of Contents

Advertisement

Installation Guide

Correcting sample vial delivery problems

Correcting sample vial delivery problems
When you find a mishandled sample vial, check:
Are there folds or wrinkles in the crimp cap, especially near the neck of the
sample vial? For more information, see your Operation Guide and your
Sampling Techniques Handbook.
Are you using the correct sample vials? Is the distance from the vial bottom
to the top of the neck about 28.4 mm? Is the bottle neck diameter
8.2 ± 0.3 mm? For more information, see your Operation Guide.
If you use labels on the sample vials, are they the correct size? For more
information, see your Operation Guide.
If you use labels on the sample vials, are they interfering with the gripper?
For more information, see your Operation Guide.
Are there any obstacles to the tray arm or injector turret motions? Remove
any obstacles.
Are the tray quadrants and turret in good condition? Clean any residue from
the sample positions. Temperature-controlled liquid pumped through the
quadrants, may cause quadrant distortion over time. See your Operation
Guide.
24

Advertisement

Table of Contents
loading

Table of Contents