Replacing The Needle Support Assembly - HP 7683 Installation Manual

Automatic liquid sampler
Table of Contents

Advertisement

Installation Guide

Replacing the needle support assembly

Replacing the needle support assembly
Use the standard needle support assembly for all injections except cool on-
column injections with a 250-µm or 320-µm column. For these injections, you
must change the needle support assembly to HP accessory 18599T.
Needle support foot
Figure 11 Needle support assemblies
To replace the needle support assembly:
1.
Remove all vials from turret, and disconnect the injector cable from the GC
and lay the HP G2613A Injector on its back on a flat surface.
2.
Open the injector door.
3.
Remove the syringe.
18
250 µm/ 320 µm
230 µm/ 320 µm
(HP accessory 18599T)
Needle support foot
530 µm
(standard)