Technical Data - E+E Elektronik HUMOR 20 Hardware Manual

Humidity calibrator
Table of Contents

Advertisement

Available languages

Available languages

T
ec
hnical Da
ta
T
ec
hnical Da
ta
General
Function principle
Working range
Inaccuracy of measurement
Inaccuracy temperature measure-
ment in measuring chamber
Power supply
Work equipment
Stabilisation time HUMOR 20
Stabilisation time specimen
Integrated power supply
Number of measuring inputs
Typ. error for display inputs
Display
Gas flow
Recommended interval for
recalibration
Interface for PC connection
System requirements for
software tools
Environmental conditions
Applied harmonised standards
Dimensions
Weight
1) The extended inaccuracy of measurement results from the standard inaccuracy increased by a multiplying factor of K=2.
2) Valid for metal covers
10

TECHNICAL DATA

two-pressure-reactor
10...95% RH
1) 2)
2)
typ. ±0.3 degC
90...230V AC
• compressed air, filtered and free of oil or nitrogen N 2 with max. 10bar
• distilled water
< 3 min/measuring point
typ. 20 min/measuring point
24V DC, max. 200mA
2 (switchable between 4...20mA / 0...20mA / 0...1V / 0...5V / 0...10V)
Voltage measuring:
Current measuring:
Dot-matrix display with backlight
3 l/min
for RH > 85% the gas flow is reduced to 1.5 l/min at 95% RH
1 year
RS232 (COM port)
MS Windows 98 / ME / NT 4.0 with SP 6a
MS Windows 2000 with SP 2 / Windows XP
temperature:
humidity:
EN 61000-6-3
EN 61000-6-2
OEVE EN 61326-1+A1+A2
400 x 260 x 240 mm
HUMOR 20: about 23kg
HUMOR 20 incl. aluminium transport case: about 36.5kg
31
relative humidity Uw [% RH]
(±0.54°F)
< 5mV
< 30µA
10...40 degC
(50...104°F)
10...80% RH
EN 61000-6-4
EN 61010-1
(15.7 x 10.2 x 9.4")
(51 lbs)
(145psi)
EN 60068-2-6
EN 60068-2-29
(80.5 lbs)
english

Advertisement

Table of Contents
loading