OPERATING ELEMENTS

5.1
HUMOR 20
6

GENERAL USER TIPS

6.1

Setup:

The humidity calibrator is designed for operation in rooms with a temperature
range of 10...40 degC
(50...104°F)
Measurements under direct sunlight as well as under the effects of other external
heat sources are not allowed.
Furthermore, the HUMOR 20 is neither to be operated in areas with explosives
nor is it to be subjected to mechanical vibrations.
A humidity calibrator that is ready for operation and that is filled with water
must not be tilted more than 20°!
6.2

Start-up:

6.2.1
Establish power supply.
Insert power supply cable into instrument and into the grounded socket.
6.2.2
Turn on main switch.
6.2.3
Turn humidity controller all the way to the left.
23
output signal of the specimen
media supply (oil-free compressed air / N2
switch between specimen 1, specimen 2,
temperature of the measuring chamber
fixing bracket for filter set
measuring input of the specimen
and a humidity content of 10...80% RH.
water intake
display
(pre-mounted)
humidity controller
powers supply +
measuring chamber
english