Page 1
H U M O R 2 0 H U M O R 2 0 FEUCHTEKALIBRATOR FEUCHTEKALIBRATOR HUMIDITY CALIBRATOR HUMIDITY CALIBRATOR CALIBRATEUR D'HUMIDITÉ CALIBRATEUR D'HUMIDITÉ Bedienungsanleitung Bedienungsanleitung Hardware Hardware Manual Manual Hardware Hardware Notice Notice Matériel Matériel BA_HUMOR20_HW_07 // Technische Änderungen vorbehalten // 302251...
Page 2
® the installation, usage and also impossibility of usage of a software product from E+E Elektronik and support services possibly associated with it or non-performance of support.
Sämtliche in dieser Anleitung enthaltenen Informationen, technische Daten und Darstellungen basieren auf zum Zeitpunkt der Erstellung verfügbaren Informationen. Die Firma E+E Elektronik ® Ges.m.b.H. behält sich das Recht vor, jederzeit und ohne Vorankündigung, Änderungen an den technischen Daten oder andere technische Modifikationen vorzunehmen, ohne eine Verpflichtung zu übernehmen, Modelle welche vor dem Datum einer solchen Änderung gefertigt wurden,...
++43 / 7235 / 605-8 info@epluse.at www.epluse.at Bezeichnung des Gerätes HUMOR 20 Der HUMOR 20 ist ein professionelles Gerät, welches eine relative Luftfeuchte darstellt, die dem Überprüfen (Kalibrieren, Justieren) von Feuchte - Messgeräten dient. Das Gerät entspricht der EG-Richtlinie Elektromagnetische Verträglichkeit (89/336/EEC)
Plexiglas-Abdeckung für Raumtransmitter-Prüfung Nordamerika (110V) O-Ring für Raumtransmitter Wasser - Ablaufschlauch mit Anschlussstück Rändelmuttern Trichter Deckel für Messkammer Bestellcode HA0202xx Inbusschlüssel (10mm) (nicht im Lieferumfang HUMOR 20 enthalten) Haltewinkel für Filterset (vormontiert) Messbecher Werksprüfzeugnis nach DIN EN 10204-3.1 Mess- und Kalibrationssoftware deutsch...
SICHERHEITSHINWEISE Allgemein Das Gerät HUMOR 20 ist nach dem heutigen Stand der Technik gebaut und somit in einwandfreiem Zustand und bei sachgemäßer Bedienung und Wartung betriebssicher. Von dem Gerät können Gefahren ausgehen, sofern es von unqualifiziertem Personal unsachgemäß oder nicht bestimmungsgemäß eingesetzt wird! •...
Wärmequellen sind nicht zulässig. Der HUMOR 20 darf nicht in explosionsgefährdeten Bereichen betrieben, noch starken mechanischen Erschütterungen ausgesetzt werden. Wird der HUMOR 20 in betriebsbereitem Zustand (mit destilliertem Wasser gefüllt) transportiert, so darf das Gehäuse nicht mehr als 20° geneigt werden! Inbetriebnahme: 6.2.1...
Page 9
6.2.4 Wasserbehälter öffnen. Verschlussschraube mit Hilfe des beiliegenden Inbusschlüssels öffnen. 6.2.5 Mit destilliertem Wasser befüllen. Den Feuchtekalibrator mit max. 1300ml. destilliertem (de - ionisierten) Wasser befüllen. Wurde max. Einfüllmenge überschritten wird Display Water - Level high angezeigt und es muss wiederum Wasser entleert werden bis die Meldung verschwindet.
40 mm bzw. 95 x 95 x 40 mm. Kubische Transmitter (Raumausführung) Unter Verwendung der im Lieferumfang enthaltenen Plexiglasabdeckung der Messkammer kann der HUMOR 20 auch für die Kalibration und Justage von Raumgeräten verwendet werden. Auf Grund der thermischen Einkopplungen von außen, ist in Abhängigkeit von der eingestellten Feuchte und Lage des Prüflings in der Messkammer mit...
Messumformer mit Stabfühler: Der HUMOR 20 hat auf Grund seines Funktionsprinzips eine geringfügig höhere Temperatur als die Umgebung. Bei der Vermessung ist daher darauf zu achten, dass sich die Fühlertemperatur an die Messkammertemperatur angleichen kann. Für präzise Vermessungen sollte die Deckeldurchführung weitestgehend dem Fühlerdurchmesser angepasst werden.
WARTUNG Das destillierte (de - ionisierte) Wasser sollte periodisch alle 8 Wochen gewechselt werden. Wird das Gerät für längere Zeit nicht verwendet, wird empfohlen das Wasser voll- ständig zu entleeren. Wasser nachfüllen (bei Störmeldung: Water - Level low Feuchte - Regler ganz nach links drehen und abwarten bis Anzeige “OUT OF SPEC”...
Leitungen. einem saugfähigen Tuch. Wasser voll- Wird meistens durch den Transport ständig entleeren und längere Zeit eines mit Wasser gefüllten HUMOR 20 Medium nachströmen lassen. Dies oder durch den Anschluss der Druckluft wird durch eine Sollwertvorgabe von bei nicht vollständig nach links gedreh- 75% über den Feuchte - Regler...
Wasser) nachgefüllt werden. 9.5 Störmeldung: heat defect Der HUMOR erreicht seine Betriebs- Kontaktieren bitte Ihren temperatur nicht. Vertriebspartner E+E Elektronik. 9.6 Störmeldung: pressure excess Es wurde ein Feuchte - Sollwert < 8% Kontaktieren bitte Ihren RH gewählt. Durch den hohen Druck Vertriebspartner E+E Elektronik.
9.8 Stabilisierungszeit zu lange Das Zeitverhalten des HUMOR 20 für Feuchtesprünge 25-80 %r.F. bzw. 80-25 %r.F. lässt sich aus dem Chart ablesen und liegt bei Sprüngen zu niedrigeren Werten bei etwa 1 Minute und bei Sprüngen zu höheren Feuchten bei etwa 3 Minuten.
400 x 260 x 240 mm Gewicht ca. 23kg (HUMOR 20) ca.36,5kg (HUMOR 20 inkl. Verpackung) 1) Angegeben ist die erweiterte Messunsicherheit, die sich aus der Standardunsicherheit durch Multiplikation mit dem Erweiterungsfaktor K = 2 ergibt. 2) gültig für Metalldeckeln...
Geräuschpegel 45 dB(A)/lm Abmessungen (L x T x H) 380 x 380 x 480 mm Gewicht 26kg Filterset mit Ölabscheider HA020109 Empfohlen bei Anschluss des HUMOR 20 an eine bestehende Druckluftversorgung. ABBILDUNG Messkammerabdeckungen FÜHLERDURCH- ANZAHL DER TYPENBEZEICHNUNG für Stabfühler MESSER [mm] DURCHFÜHRUNGEN...
Page 18
5.1 HUMOR 20 GENERAL USER TIPS 6.1 Setup 6.2 Start-up CALIBRATION AND ADJUSTMENT PROCESS HUMOR 20 7.1 Compact room transmitter 7.2 Transmitter with sensor probe 7.3 Interruption of power supply during the measurement process 7.4 Ending the calibration or adjustment process MAINTENANCE 8.1 Adding water (for fault message: Water - Level low)
Furthermore, the manufacturer is convinced that the humidity calibrator HUMOR 20 meets all requirements and expectations that you might have as the buyer of a new instrument. With proper handling and regular maintenance, the instrument should operate reliably for many years.
Designation of the Instrument HUMOR 20 The HUMOR 20 is a professional instrument that describes relative air humidity used to test (calibrate, adjust) humidity instruments. The instrument corresponds to EG guideline Electromagnetic Compatibility (89/336/EWG) EG guideline Low Voltage (73/23/EWG)
Water drain pipe with connector Knurled nut Cover für measuring chamber, ordering code HA0202xx Funnel Allen key (10mm / (not inlcuded in the scope of supply HUMOR 20) 0.4” Measuring beaker Fixing bracket for filter set (pre-mounted) Works certificate acc. DIN EN 10204-3.1...
SAFETY INSTRUCTIONS General The instrument HUMOR 20 is built according to the current state of the art and it will operate reliably if it is free from faults and if it is properly operated and maintained. The instrument can present a risk if it is used improperly or not according to specifications by unqualified personnel! •...
Measurements under direct sunlight as well as under the effects of other external heat sources are not allowed. Furthermore, the HUMOR 20 is neither to be operated in areas with explosives nor is it to be subjected to mechanical vibrations.
Page 24
6.2.4 Open water container. Open screw plug with the help of the included Allen key. 6.2.5 Fill with distilled water. Fill the humidity calibrator with max. 1300 ml distilled (deionized) water. If the max. fill lever is exceeded, Water - Level high is displayed on the display and water must be removed until the message disappears.
Compact room transmitter With the use of the Plexiglas cover of the measuring chamber included in the scope of supply, the HUMOR 20 can also be used for the calibration and adjustment of room devices. Due to thermal coupling with the outside, additional measurement errors are to be considered depending on the used humidity and the position of the specimen in the measuring chamber.
Transmitter with sensor probe: Due to its working principle, the HUMOR 20 has a slightly higher temperature than the surroundings. Therefore, for the measurement, make sure that the probe temperature can be aligned with the measuring chamber temperature. For precise measurements, the cover-feedthrough should be adapted to the probe diameter as much as possible.
MAINTENANCE The distilled (deionized) water should be changed periodically every 8 weeks. If the instrument is not used for long periods, it is recommended to completely remove the water. Adding water (for fault message: Water - Level low) Turn humidity - controller all the way to the left and wait until the display indicates “OUT OF SPEC”.
WARNINGS / ERROR MESSAGE ON THE DISPLAY Line: Display Humidity () or Error () Line: Status line () or at HUMOR 20 current measuring value of the rel. HUM. =100.0% selected specimen WATERLEVEL LOW rel. HUM. =100.0% Your humidity calibrator is equipped with a self-diagnosis system. According to...
Turn the controller all the way to the left close compressed air. After a stabilization time of 2h, the display should show 100 ± 2% RH. If this is not the case, please contact your sales partner E+E Elektronik. english...
Stabilization time too long The chart shows the stabilisation time of the HUMOR 20 at a humidity step from 25 to 80% RH and 80 to 25% RH. The stabilisation time at low humidity is approx. 1 min., at high humidity approx. 3 min.
Weight HUMOR 20: about 23kg (51 lbs) HUMOR 20 incl. aluminium transport case: about 36.5kg (80.5 lbs) 1) The extended inaccuracy of measurement results from the standard inaccuracy increased by a multiplying factor of K=2. 2) Valid for metal covers...
(15 x 15 x 18,875 “) Weight 26kg (57lbs) Filter set with oil separator HA020109 The filter set is recommended when connecting the HUMOR 20 to an existing compressed air system. Optional covers for the PROBE NUMBER OF TYPE PICTURE...
Page 33
MATERIEL AVANT PROPOS DECLARATION DE CONFORMITE CE ELEMENTS DE LA FOURNITURE INSTRUCTIONS DE SECURITE 4.1 Généralités 4.2 HUMOR 20 ELEMENTS FONCTIONNELS 5.1 HUMOR 20 CONSEILS GENERAUX DESTINES A L'UTILISATEUR 6.1 Installation 6.2 Mise en marche PROCEDURES DE CALIBRATION ET D'AJUSTEMENT DU HUMOR 20 7.1 Transmetteur compact de laboratoire...
En outre, le fabricant est convaincu que le générateur d'humidité HUMOR 20 satis- fait tous les besoins et toutes les attentes que vous pouvez manifester en tant qu'acheteur d'un nouvel instrument.
Désignation de l'appareil HUMOR 20 Le HUMOR 20 est un instrument professionnel pour la génération de l'humidité relative de l'air utilisé pour étalonner (calibrer, régler) les instruments de mesure d'humidité. L'instrument correspond à : Instruction EG relative à la Compatibilité électromagnétique (89/336/EWG) Instruction EG relative aux applications basse tension (73/23/EWG) Normes harmonisées appliquées...
Entonnoir Ecrou moleté Couvercle pour chambre de mesure référence HA0202xx Clé Allen (10 mm) (non inclus dans la fourniture du HUMOR 20) Bécher gradué Logiciel de mesure et de calibration Equerre de fixation du filtre (pré-montage) Certificat de fonctionnement conforme à la norme DIN EN 10204-3.1...
INSTRUCTIONS DE SECURITE Généralités L'instrument HUMOR 20 est construit conformément aux règles de l'art actuelles. Il fonctionnera de manière fiable à condition d'être utilisé et entretenu correctement. L'instrument peut présenter un risque en cas d'utilisation incorrecte ou non confor- me aux spécifications par du personnel insuffisamment qualifié ! •...
Toute mesure soumise au rayonnement direct du soleil ainsi qu'aux effets d'autres sources de chaleur n'est pas admise. En outre, le HUMOR 20 ne doit pas être utilisé en zones explosives et ne doit pas être soumis à des vibrations mécaniques.
Page 39
6.2.4 Ouvrir le récipient d'eau. Dévisser le bouchon à l'aide de la clé Allen fournie. 6.2.5 Remplir d'eau distillée Remplir le calibrateur d'humidité avec 1300 ml maxi. d'eau (désionisée) distillée. Si le niveau de remplissage maxi. est dépassé, le message d'alarme "Water - Level high"...
Transmetteur compact de laboratoire A l'aide du couvercle en Plexiglas de la chambre de mesure compris dans le cadre de la fourniture, le HUMOR 20 peut également être utilisé pour la calibration et l'ajustement de dispositifs de mesure compact. En raison du couplage thermique avec l'extérieur, des erreurs de mesure addition- nelles doivent être prises en compte selon l'humidité...
Transmetteur avec sonde de mesure : En raison de son principe de fonctionnement, le HUMOR 20 a une température légèrement plus élevée que son environnement direct. Par conséquent, il est nécessaire de vérifier que la température de la sonde soit en équilibre avec la température de la chambre de mesure en vue de la calibration.
MAINTENANCE L'eau distillée (désionisée) devra être changée périodiquement toutes les 8 semaines. Si l'instrument n'est pas utilisé durant une longue période, il est recommandé de vider complètement le réservoir d'eau. Ajout d'eau (en cas de message d'erreur : Water - Level low (Eau - Niveau bas) Tourner la molette de réglaged'humidité...
Ligne : Affiche l'Humidité () ou les Erreurs () Ligne : Ligne d'état () ou valeur de mesure actuelle au niveau rel. HUM. =100.0% du HUMOR 20 correspondant au spéci- WATERLEVEL LOW sélectionné. rel. HUM. =100.0% Votre calibrateur d'humidité est équipé d'un système d'autodiagnostic. Selon l'état de fonctionnement ou les erreurs survenues, différents messages de service ou...
Le HUMOR n'atteint pas sa tempéra- Merci de bien vouloir contacter le ture de service. réseau de vente E+E Elektronik. 9.6 Message d'erreur : pressure excess (pression excessive) Une valeur d'humidité souhaitée < 8% Merci de bien vouloir contacter le d'HR a été...
4. Retirer la carte électronique (voir l'illustration 3). 5. Mettre en place la carte électroni- que de rechange et fermer le boîtier. 6. Calibrer le générateur HUMOR 20. l'illustration La calibration usine disparaît lors du remplacement d'une carte électronique ! Pour toute question, merci de contacter le réseau de vente E+E Elektronik.
Equipement d'installation • air comprimé, filtré déshuilé ou azote à 10 bars de pression max. • eau distillée Durée de stabilisation HUMOR 20 < 3 min/point de mesure Durée de stabilisation des instruments typique 20 min/point de mesure Alimentation électrique intégrée 24V DC, max.
Certificat d’étalonnage OEKD20/xH Pour satisfaire aux exigences des Systèmes d’Assurance Qualité tel que ISO9001 concernant la calibration et les certificats de tests et de mesure des instruments, le HUMOR 20 est disponible avec un certificat d’étalonnage officiel accredité OEKD. français...