CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION

Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing to be cleaned
and disinfected.
Cleaning
Important: In order to execute cleaning or any type of maintenance, ensure that the equipment is
disconnected from the electrical network.
Attention
Never wet the compressor or any of their parts. The cleaning of outside surfaces should
be performed weekly with a duster or soft cloth slightly moistened with water; the use of
other products may damage the compressor finish or plastic parts. Be sure to thoroughly
clean the head's upper part, responsible for its cooling. The accumulated dust affects the
heat dissipation.
Preventive Maintenance
The equipment should be calibrated routinely, as per the legislation in force in the country.
But never with a period exceeding 3 years.
In order to protect your equipment, seek Alliage technical assistance for periodic revisions of
preventive maintenance.
Filter Change:
Keep the air filter(s) clean and in good condition of usage. The air filter prevents dust from entering
into the compressor and reduces the operating noise.
Over time, the filter will become dirty, reducing the air inlet flow and causing the decrease of
compressor performance. The filter should be replaced every 3 months on average or before,
depending on the air quality. Additional filters are available for purchase through the Alliage
Technical Assistance network.
To perform the change, open the filter enclosure by turning the metal cover; remove the old filter and
insert the new one; put the cover again by turning it in the opposite direction.