ENGLISH Aerosol Mistral is a system for aerosol therapy, with 230V ~ / 50Hz power supply, intended for household use. The appliance is designed for non-continuous use (Ton: 20 mins / Toff: 40 mins). The medical device is designed to be easy to transport and use and is indicated for nebu- lizing bronchodilators and antibiotics.
Page 3
ENGLISH on the adapters and on the extension cables. 5. For repair works contact only Gima’s technical service or a technical center authorized by the manufacturer. The device requires the use of original spare parts. Failure to com- ply with the above may compromise the safety of the appliance.
In any event, the temper- atures do not exceed the limit of 105°C [221°F] (ref. Interpretation Sheet IEC 60601-1). Gima S.p.A. cannot be held liable for accidental or indirect damage resulting from alterations of the device, repairs and/or unauthorized technical interventions, or damage to any of its part due to accident, misuse and/or abuse.
Percentage of humidity in the environment: 0 ÷ 93% RH Atmospheric pressure: 500 ÷ 1060 hPa Copy of the EC Declaration of conformity can be requested to Gima S.p.A. - Via Marconi, 1 - 20060 – Gessate, Milan (Italy) MMAD = Mass Median Aerodynamic Diameter...
ENGLISH MAINTENANCE The Mistral device does not have any part that requires maintenance and/or lubrication. Nonetheless, it is necessary to carry out some checks to verify the function and security of the appliance before each use. Remove the appliance from the box and always check that there is no visible damage;...
ENGLISH PREPARATION 1. Pull out the air hose from the nebulizer leaving it plugged to the air outlet pipe of the device. 2. Rotate the upper part of the nebulizer anticlockwise. 3. Disconnect the internal pisper at the bottom of the nebulizer with finger force. CLEANING Clean all the components of the nebulizer (except the air hose) before and after each use by choosing one of the methods described below.
Page 8
ENGLISH 1 – Air hose 2 – Ampoule lower part 3 – Nebulizer Nozzle 4 – Ampoule upper part 5 – Mouthpiece 6 – Adult Mask 7 – Children Mask 8 – Nosepiece (non-invasive) Use only the original accessories intended and indicated by the Manufacturer.
Page 9
It is recommended to completely unroll the power cable to avoid dangerous overheating. In the event of damage to the power cable, contact Gima’s technical service for the replacement. 2. Open the nebulizer by unscrewing the cover anticlockwise.
Page 10
Guide and declaration of the manufacturer – Electromagnetic Emissions The Mistral aerosol can be used in the following electromagnetic environment. The Cus- tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in such environment.
Page 11
Guide and declaration of the manufacturer – Electromagnetic Immunity The Mistral aerosol can be used in the following electromagnetic environment. The Cus- tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in such environment.
Page 12
The Mistral aerosol is designed to operate in an electromagnetic environment in which the RF radiated interferences are kept under control. The customer or the user of the Mistral appliance can contribute to prevent electromagnetic interferences by ensuring a minimum...
Page 13
The warranty is valid for 12 months from the date of supply of GIMA. During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected parts due to production reasons. Labor costs and personnel traveling expenses and packaging not included.