44 |
9
Troubleshooting
Messages may occur during system operation due to operator error or component failure. For further assistance, contact SCP SCIENCE
Service Department or your local distributor.
List of Error Messages:
No.
Error Message
1
COM.ERROR
2
ERROR! ID=Stalled DESC= Section 1 (2,3,4) is stalled. FA
3
ERROR!!! ID= Timed Out. DESC= Section Operation has timed out. FA
4
ERROR COM Timed out! FA
5
ERROR No OK received from MB. FA
6
Process terminated with exit code #. Press OK to restart.
7
DIGESTION.RUN TUBE 01 E_CODE=02
8
DIGESTION.RUN TUBE 01 E_CODE=03
9
DIGESTION.RUN TUBE 01 E_CODE=04
10
MB_SYNTH
11
ERROR Digestion Operation has timed out!
12
ERROR DIGESTION_ALARM_ERROR Magnetron is alarmed.
13
DIGESTION_OVERHEAT_ERROR Magnetron overheated
14
ACTUATOR_STALLED_ERROR
15
ACTUATOR_OPERATION_TIMEDOUT_ERROR
16
CAVITY_STALLED_ERROR
17
CAVITY_UNKNOWN_RACK
18
CAVITY_OPERATION_TIMEDOUT_ERROR
19
CAVITY_OBSTRUCTED_ERROR
20
CAVITY_NOT_INITAILIZED_ERROR
21
CAVITY_NOT_NOW_ERROR
22
CAVITY_ORITENTATION_FAILED_ERROR
23
CAVITY_INTERNAL_ERROR
24
RACK ID FAILED
25
RACK IN REVERSE POSITION
26
PMB
Version 0.15.1
Description
Unrecognized packet.
Rack is stalled at B-F section.
Operation B-F timed out. (R-L, F-B or L-R).
No response from MB.
OK not received from MB.
GUI crashed code number #. Press OK to resume.
Never reached target temperature.
Temperature rising too slowly in tube number #.
Tube number # overheats.
System initialize.
Digestion operation has timed out on tube number #.
Magnetron number # has an alarm.
Magnetron number # overheats.
Actuator is stalled.
Actuator operation has timed out.
Tunnel motor is stalled.
Please remove unknown rack from tunnel.
Tunnel operation has timed out.
Tunnel is obstructed.
Tunnel motors need initialization.
This command cannot be performed.
Unknown rack orientation.
An error occurred in tunnel.
Thermal switch has failed.
Need help?