5
Error Information

Thermostat Error Messages

Thermostat Error Messages

Error messages are displayed in the user interface when an electronic failure
occurs with the autosampler thermostat module which requires attention. In
the event of such a failure, the red status indicator at the front of the
thermostatted autosampler is switched on, and an entry is written into the
instrument log book.
This section describes the meaning of autosampler thermostat module error
messages, and provides information on probable causes and suggested actions
how to recover from error conditions.

Fan Failed

Each fan in the thermostat is equipped with a speed sensor, that allows
monitoring and control the speed of the fans. If the sensor shows no signal,
when the fan is activated, it is very likely that the fan is defective. The number
in brackets indicates the fan position. Position numbers of the fans are as
follows (seen from front of the thermostat)
• fan failed 1 - left side, fan in front position
• fan failed 2 - left side, fan in back position
• fan failed 3 - right side, fan in back position
• fan failed 4 - right side, fan in front position
• fan failed 5 - small cooler fan on top of cooling / heating module
Probable cause
1
2
52
One of the 4 heat-sink fans is defective.
The cooling heating module fan is defective
Suggested actions
See Service Manual for Exchanging the
Heatsink Fans.
See Service Manual for Exchanging Heat
Exchanger Fan.