Download Print this page

TESTO 0554 5762 Application Information page 2

Probe module multi-hole probe

Advertisement

Available languages

Available languages

Application information
Please read this documentation through carefully and familiarise yourself with the operation of the
product before putting it to use. Keep this document to hand so that you can refer to it when
necessary. Please refer to the Instruction manual on the analyser.
Intended use
flue gas probes with a diameter of 60 to 300 mm (0554 5762) or 60 to 180mm (0554 5763).
The whole cross-section is measured up to a gas pipe diameter of 200 mm. In the case of
measurements in flue gas pipes with a diameter of more than 200 mm, markings guarantee that the
measuring field is located in the middle of the flue gas pipe (0554 5762 only).
Operation
In order to avoid damage, the probe pipe lock must be released before the probe pipe is pulled
out or pushed back again.
1 Release probe pipe lock by turning the probe stop.
2 Turn the probe pipe so that the openings point against the flue gas flow. For this purpose pull out
the probe pipe as far as it will go, turn it into the required position and push it back again.
3 Flue g g as p p ipe d d iameter u u p t t o 2 2 00mm:
Pull the probe pipe a little further out than the diameter of the flue gas pipe to be measured.
Flue g g as p p ipe d d iameter m m ore t t han 2 2 00mm ( ( 0554 5762 o o nly):
Pull out the probe pipe until the tip of the probe holder corresponds to the respective marking:
· 1. Marking: 200 mm
· 2. Marking: 250 mm
· 3. Marking: 300 mm
4 Close probe pipe lock by turning the probe stop, tighten probe stop somewhat.
5 Insert multi-hole probe as far as probe stop in the measurement opening and measure.
Hot metal parts!
Can c c ause b b urns.
Let probe stop and probe pipe cool down after the measurement before touching
them.
Probe module Multi-hole probe 0554 5762, 0554 5763
Measurement of CO/CO
undiluted CO level in the flue gas of gas furnaces
The multi-hole probe can be used in universal
applications and has a continuous working range in
/O
mean value to determine
2
2
en

Advertisement

loading

This manual is also suitable for:

0554 5763