Biosan ES-20/60 Operating Manual

Biosan ES-20/60 Operating Manual

Orbital shaker-incubator

Advertisement

Quick Links

Advertisement

loading
Need help?

Need help?

Do you have a question about the ES-20/60 and is the answer not in the manual?

Questions and answers

Summary of Contents for Biosan ES-20/60

  • Page 2: Table Of Contents

    Contents Safety precautions ....................3 General Information ....................4 Getting started ......................4 Operation ......................... 6 Specifications ......................8 Maintenance ......................9 Warranty and Claims ....................10 EU Declaration of Conformity ................. 11...
  • Page 3: Safety Precautions

    Safety precautions The following symbols mean: Caution! Make sure you have fully read and understood the present Manual before using the equipment. Please pay special attention to sections marked by this symbol. GENERAL SAFETY  Use only as specified in the Operating Manual provided. ...
  • Page 4: General Information

    It provides reliable and stable operation for the long term experiments needed for cell cultivation. ES-20/60 Orbital Shaker-incubator provides smooth (or intensive) mixing in flasks installed on the platform. Built-in thermoresistant brushless fan provides precise temperature distribution inside the chamber (from 10°C above ambient up to +80°C).
  • Page 5       3.3. Setup: place the unit upon even horizontal non-flammable surface at least 30 cm away from any flammable materials; Note! Ensure that the unit is placed on solid, level surface, which is able to support its weight.
  • Page 6: Operation

    Operation Fig.1. Control panel 4.1. Connect the unit to a grounded power socket. Set the Power switch on the front side to position I (ON). 4.2. The display will turn on with the upper line showing the previously set time, speed and temperature and the lower line showing current readings of the same parameters (chamber temperature °C, which automatically starts rising according to the temper- ature set in the upper line).
  • Page 7 4.4. Program execution. 4.4.1. Open the door and place the samples on the platform. Close the door. Caution! Please make sure that all vessels are tightly sealed. Humidity caused by evaporation from unsealed vessels inside the incubator will dam- age the unit! 4.4.2.
  • Page 8: Specifications

    Clamp for 1000 ml flask for UP-168 platform (ø 130 mm) FC-1000 BS-010126-IK Biosan is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice. Accurate within ±10%.
  • Page 9: Maintenance

    Maintenance 6.1. If the unit requires maintenance, disconnect the unit from the electric circuit and con- tact Biosan or your local Biosan representative. 6.2. All maintenance and repair operations must be performed only by qualified and spe- cially trained personnel.
  • Page 10: Warranty And Claims

    7.5. Extended warranty. For ES-20/60, a Premium class model, one year of extended warranty is available free of charge after registration, during 6 months from the date of sale. Online registration form can be found in section Warranty registration on our website at the link below.
  • Page 11: Eu Declaration Of Conformity

    EU Declaration of Conformity Unit type Shakers-incubators Models ES-20, ES-20/60 Serial number 14 digits styled XXXXXXYYMMZZZZ, where XXXXXX is model code, YY and MM – year and month of production, ZZZZ – unit number. Manufacturer SIA BIOSAN Latvia, LV-1067, Riga, Ratsupites str. 7/2...