Response; Power Supply; Isolation - Knick SensoTrans A 20220P0 Series Manual

Table of Contents

Advertisement

Available languages

Available languages

Response

Characteristic
Measuring rate
Response time t 99 *
* Time after change of input value until reaching an output value of 99 % steady
state

Power Supply

Power supply

Isolation

Test voltage
Working voltage
(basic insulation)
Rising / falling linearly
Approx. 3/s
300 ms
24 V DC (-20 %, +25 %)
approx. 0.85 W
2.5 kV, 50 Hz: power supply against
input against output
Up to 300 V AC/DC across all circuits
with overvoltage category II and
pollution degree 2. For applications
with high working voltages, take
measures to prevent accidental
contact and make sure that there
is suffi cient distance or insulation
between adjacent devices.
Specifi cations
47