Non-Contact Temperature Measurement - mru 400DH User Manual

Hand detector probe
Table of Contents

Advertisement

12 / 14
USER MANUAL 400DH

4.4. Non-contact temperature measurement

The IR400 interchangeable sensor is used for non-contact temperature measure-
ment.
Aim the sensor at the surface to be measured. The measuring distance depends on
the size of the surface to be measured, the closer the measurement becomes, the
more accurate it becomes. The minimum distance is approx. 1 - 2 cm.
Functions in the Context menu:
function
Signal ON (Key F1)
Signal ON (Key F1)
Max = 0 (Key F3)
Reset
Setting
Printout (Dr)
Meas. menu (ESC)
Info
Settings:
Signal threshold
units
Technical Specification:
range
Sensor IR400
-70 ... +380 °C
explanation
With this function the signal is switched on.
With this function the signal is switched off.
The value Max is set to zero.
This function resets the graphic.
Setting units and signal thresholds
Printout is generated, a printer is required for this
function.
Exit measuring program
The information (sensor name, page number, firmware
version) about the sensor is displayed here.
30 °C ... 380 °C
° C, ° F
resolution
0,1 °C
field of vision
35°
MRU GmbH, D-74172 Neckarsulm