Micro Direct MicroLab Operating Manual page 19

Table of Contents

Advertisement

The calibration error for expiration
and inspiration at each flow rate
are displayed.
The calibration
error should be less than 3.5%. If
a greater error is shown, repeat
the procedure ensuring that the
syringe is emptied and filled in a
smooth manner without jerking the
handle.
If an error greater than
3.5% is still shown, inspect the
turbine transducer and clean if
necessary.