III. Calibrating the Data Logger Option
Connect the computer to the USB port on the FC220 or PC320 data logger.
Launch the Global Logger II software and connect to the logger, you must use
Global Logger II version 2.1.8 or higher. Click the Setup button to enter the
calibration screens. A few settings are required for the logger option to operate
properly, these are shown below. In the General tab; set the Sensor Warm-Up
Time to zero seconds and enable Sample-On-Demand by checking the checkbox.
The logger will not record relay events unless the Sample On Demand function is
enabled.
Select the Analog Channels tab.