Appendix 5: Powder Dispense Parameter Step - RCBS MATCHMASTER 98941 Product Instructions

Powder scale/dispenser
Hide thumbs Also See for MATCHMASTER 98941:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Appendix 5: POWDER DISPENSE PARAMETER STEP FUNCTION
This feature allows the user the ability to step through the user defined powder dispense
parameters, E thru I.
1.
Upon completion of setting Powder Dispense Parameters, A-I; depress "GO" to save, "Test"
will display.
2.
If you choose not to run test and step through parameters, E-I; depress the "Cancel"
button to exit out of Test and begin dispensing.
3.
If a test is desired on the recently set Powder Dispense Parameters depress "GO" to begin
step operation.
4.
Enter target dispense weight and depress "GO".
5.
Unit will dispense until it reaches parameter "E", dispensing will stop and the display will
flash between the dispensed weight and Parameter "E". If unit finishes dispensing and
displays Parameter "A", the parameters entered caused an overcharge condition. When
this happens the unit will default back into user configuration and the parameters will
need to be modified. Make sure parameter "A" represents the size of powder being used.
6.
If parameter "E" does not need to be changed, press "GO" to accept and continue to
parameter "F".
7.
If parameter "E" needs to be changed, depress the "CAL" button. Parameter "E" will display,
the value to the right of E will be changed, the value to the left of E will not change as this
is the recommended setting and is displayed only as a reference. Enter desired value and
press "GO" to accept.
8.
After parameter has been changed:
Empty scale pan
Enter charge weight
Depress "GO" to dispense.
9.
Unit will dispense until it reaches parameter "E". If value is acceptable depress "GO" to
continue Step process to parameter "F", if not repeat steps 7-8 above.
10. Unit will dispense until it reaches parameter "F".
If value is accepted depress "GO" to continue Step process and test parameter "G".
If value needs to be modified follow the same process as described in steps 7-8 for
parameter "E".
Note: when step process is run it will start over and step through previous parameters, i.e.,
dispense until parameter "E" and stop, depress "GO" to accept and continue to parameter "F"
11. If Parameter "F" has been accepted and "GO" is depressed, the Long Trickle operation,
parameter "G" will begin and trickle until Parameter "H" is reached.
Display will flash between weight and parameter "G".
If "G" is acceptable depress "GO" to display parameter "H"
If parameter "G" needs to be changed depress the "CAL" button. Parameter "G" will
display. Change the value and select "GO" to accept.
Empty Scale pan
Enter charge weight
Depress "GO" to dispense
Unit will step through Parameters, E-G, depress "GO" at each step to accept; if
parameters produce desired results. If not, depress "CAL" at any step to modify value.
19

Advertisement

Table of Contents
loading

Table of Contents