C C o o m m p p u u t t e e r r S S e e t t - - U U p p
The computer's set-up sequence is entered automatically when the batteries are changed. You can also enter
the set-up sequence at any time by holding both buttons simultaneously for 8 seconds. During the set-up
sequence, all of the display segments are displayed first, and then "mi" appears for English units and "km"
appears for metric units. The left button toggles between English and Metric units.
If you wish to measure in English units (standard), or in "feet per second", toggle to "mi". Push the right
button to enter "CAL" mode. This is your Flow Probe calibration function. Set the calibration at 3 3 . 3 1.
When you change your batteries, you must reset this number. Pushing the left button increases the number
when the arrow points up and decreases the number when the arrow points down.
If you wish to measure in Metric units, or in"meters per second", toggle to "km". Push the right button to
enter "CAL" mode. This is your Flow Probe calibration function. Set the calibration at 1 6 0 3. When you
change your batteries, you must reset this number. Pushing the left button increases the number when the
arrow points up and decreases the number when the arrow points down. NOTE: Unless Metric units are
specified when ordering, then measuring in metric units (meters per second) the numbers on the display read
10 times higher than actual measurements (2.23 meters per second reads 22.3).
To continue the set-up sequence after you have set your English or Metric calibration:
• Push the Right button - be sure "CAD" is not displayed.
• Push the Right button - SLEEP will appear. If you are not using your Flow Probe for 1-2 month, leave it
in this SLEEP mode, to reduce battery drain.
• Push the Right button - push the Left button to toggle between 24 hr and 12 hr clock.
• Push the Right button - push the Left button to set HOUR (time of day).
• Push the Right button - push Left to set the MINUTE (time of day).
• Push the Right button - you are now out of Set Up and back in Velocity ("V").
FP101/FP201 - 4
Global Water Instrumentation, Inc.
www.globalw.com • (800) 876-1172 • US (916) 638-3429 • FAX: (916) 638-3270 • E-mail: globalw@globalw.com