If a problem persists, call your supplier or Tamson Instruments bv. WARRANTY Tamson Instruments bv. warrants that all their manufactured equipment is free from defects in material and workmanship, preventing the machine from normal operation. Tamson Instruments bv does not warranty that the equipment is fit for any other use than stated in this manual.
INSTALLATION Important Tamson Instruments bv is not responsible for any consequential damage or harm caused by using this bath. Repairs on the electrical system of the bath may only be carried out by well trained and authorized persons.
When removing bath fluid do not inhale toxic vapor. Always use appropriate ventilation. Handle old bath fluid as toxic waste. Cool down bath fluid to ambient before removing. Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
Low fluid level Can cause fire when heater get's partially exposed Will damage the heater Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
A lower level than 5 cm below the lid may damage the heaters. A high bath level can cause overflow and might also damage the bath insulation. Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
For < 10ºC THAN example, in the volume proportion of 50 / 50%. ADD ETHYLNE- GLYCOL Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
Pay attention when removing hot fluid: Use heat protective clothing and wear safety glasses. WEAR HEAT PROTECTIVE CLOTHES Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
Page 11
Water and oil must at all times be kept separate within the bath. Never mix oil and water in or around the bath. Do not spill water in hot oil. Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
Sufficient cooling power is applied when the indicated power exceeds 8% on the display. The bath will than be able to realise stable temperature control. Cooling circulator TLC10-3 Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
Use a mains supply that is well earthed, clean of interference and suitable for the acquired electrical load of the bath. PROTECTIVE EARTH Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
0.01°C (0.001°C for the TV4000MKIIDC) and a specific offset of 0.005°C is available from the menu. This Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
’errors’. In this way there is a continual check of the proper functioning of the bath. Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
Place the bath spirit level. The 4 supporting feet can Turn feet to be turned in and outwards for exact adjustment. adjust bath spirit level Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
Temperature measured in calibration block Measurements have been performed with silicone oil <3 mm TV4000MKII DC stability @ set point temperature Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
P + I control 2,2xB 0,8xT P + I + D control 1,7xB 0,5xT 0,12xT Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.: NL28 INGB 0007 350 370 Van ’t Hoffstraat 12...
·The operating temperature is around or below 30°C. To stabilize the bath at a set-point temperature just above Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
$00 : control algorithm disabled – allows external PC to take control of heaters $10 : startup hold-off started $11 : startup holdoff in progress Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
Page 28
$40 : Start of ratio process $41 : Ratio process in progress $42 : waiting for PV<SP-tuneoffset after ratio $50 : Start of successive approximation process Tamson Instruments bv ISO 9001 : 2015 NL/PRO 238239125 VAT: NL 80 66 34 984 B01 Bank account no.:...
Following equipment is in complience with EMC Directive 2014/30/EU: Product: Thermostatic bath Model: TV2000-MKII, TV4000-MKII, TV4000-MKII DC, TV17000 Serial code: Effective from 08Txxx Manufacturer: Tamson Instruments bv van 't Hoffstraat 12 2665 JL Bleiswijk The Netherlands The products are in conformity with the following specifications: Item Reference Description...
The equipment conforms with all the specifications and norms in this regard. The equipment conforms without any further notice. Entity responsible for marking this declaration : Manufacturer, Tamson Instruments bv, van 't Hoffstraat 12, Bleiswijk The Netherlands, Name R.C. van Hall...
Need help?
Do you have a question about the PMT TV2000MKII and is the answer not in the manual?