Error_Xxx - H2Scan HY-ALERTA 1600 Operating Manual

Intrinsically safe area hydrogen monitor
Table of Contents

Advertisement

13.5 ERROR_XXX

The following table lists possible errors, the code numbers that could appear on the display, and
their binary equivalents. These codes are hexadecimal numbers representing eight bits with each
bit representing an error. If more than one error occurs concurrently, their values will add. For
example if there is an error calculating hydrogen and the H2 resistor is out of range, the error code
will be 90. Many of these errors could lead to a "htroff" error.
90000012 R8
Error
Error calculating hydrogen
PCB temperature is too high
Sensor temperature is out of range
H2 resistor value is out of range
Configuration error
HY-ALERTA™ 1600, Intrinsically Safe Area Hydrogen Monitor
Code (hex)
80
40
20
10
04