en - Areas of application
38
For validation purposes please make sure that you follow the
loading instructions given on the template.
 To minimise the risk of injury, the mobile unit should be loaded
from the back forwards and unloaded in the reverse order.
 Any hollow items that have become disconnected from their in-
jector nozzles, adapters or holders during processing must be re-
processed.
Before loading the machine and before starting a
programme
– Check that fittings such as irrigation sleeves and injector nozzles
are securely screwed into position.
 Make sure that all screw connectors are fitted with injector
nozzles, adapters, irrigation sleeves or blind stoppers to ensure
that all rinsing fittings in use are supplied with sufficient standard-
ised pressure.
Damaged fittings such as injector nozzles, adapters and irrigation
sleeves must not be used.
Rinsing positions not in use do not have to be fitted with blind
stoppers.
– Check that spray arms are rotating freely.
– Check spray arm jets for blockages.
– Check that the basket is correctly connected to the water supply in
the washer-disinfector.