Carefully study and observe all items of these operating instruc- tions before commissioning the unit. In principle, the Wöhler IR Temp 210 should only be used by skilled personnel for its intended purpose and within the speci- fied range of data. Any liability or warranty for results established or for defects caused when using the unit is excluded by all means.
Page 3
1. Specifikation 1.2 Technical Data: Measuring Range: - 40 to + 500 °C (-40 to 932 °F) Accuracy: ± 2 % of reading or ± 2 °C Resolution: 0,1 °C /°F Distance/Spot Ratio: 10:1 Response Time: 1 sec Power Consumption: 12 mA Max.
3. Operation 3. Operation Temperature measuring The Wöhler IR Temp 210 has a laser pointer beam to mark the measuring area on the object. Turn the instrument on by pressing the blue trigger. If you only press the trig- ger for a short moment, the display will show the actual value. After approxi- mately 10 seconds the meter will turn off automatically.
4. Maintenance 3.2 Battery Replacement The instrument is powered by a 9V battery. A flashing display indicates that the battery voltage has fallen into the critical region (6.5 to 7.5 V). 4. Maintenance Clean the exterior housing with a damp cloth. Ensure that no water pene- trates into the instrument or has contact with the infrared lens.
This declaration is based on third party measurements. Guarantee and Service 8.1 Guarantee The guarantee period for the Wöhler IR Temp 210 is from the sales date, provided that it is used correctly. We see as a very important element in our business.