6.3. Application of the apparatus
6.3.1. To change the therapy mode shortly press the
button.
To set or change the electrostimulation power use
and
Attention!
The power increase is controlled subjectively follow-
ing the patient's sensations upon contact of the elec-
trodes with the skin surface. Do not surpass the pain
threshold.
At the power level more than zero and with the contact
of all electrodes with the skin surface, the apparatus
will switch to the mode of electrostimulation in com-
pliance with the selected mode and power level.
If in the process of therapy at least one of the electrodes
looses contact with the skin surface, you will hear a
sound signal corresponding to this event and the pro-
cedure will stop for 3 seconds. If during this time con-
tact with skin is restored, the procedure will resume;
in case the contact is not restored, the apparatus will
switch into the stand-by mode.
During the procedure it is allowed changing the inten-
sity of the electric stimulating impulse and the proce-
dure is neither paused nor resumed.
buttons.